Creating a New Wash Protocol
To specify a Fill (within a Move):
Note
: A
Fill
must be performed within a
Move
cycle.
1. Scroll right to display the next selection. The display will read:
Select option
SOAK FILL
2. Select FILL. The display will read:
Fluid volume? 50
3. Use the keypad to enter the volume of fluid to be dispensed
during well filling.
Note
: A volume ranging from 50 to 300
microliters can be entered.
4. Press ENTER. The display will read:
Select option
SOAK FILL
52
Opsys
MW™
User’s Guide