OPERATOR’S MANUAL
GAC® 2500 Models (UGMA & INTL)
11001-1505E-201305
MAINTENANCE / 77
CLEANING THE TEMPERATURE SENSOR PROBE
The temperature sensor may require cleaning due to dust buildup and/or
foreign material that has collected around the sensor that could potentially
cause temperature error readings during analysis. It is important the
temperature sensor is visually inspected and cleaned each time a regular
maintenance check of the instrument is conducted.
TEMPERATURE SENSOR LOCATION
With the instrument placed on its backside and looking inside the
instrument, the sensor probe is located in the middle of the instrument
toward the top and front attached to a circuit board.
Figure 72
Sensor Probe Location
CLEANING METHOD
The temperature sensor probe is hermetically sealed so the best cleaning
method is using
oil-less
compressed air with an extension tube to clear
particles from around the sensor.
1. Unplug the power cord and all cables.
2. Remove grain drawer.
3. Turn the instrument on its backside (
Figure 72
).
4. Hold the can upright pointing the extension tube approximately 2
inches from the sensor probe thoroughly cleaning around the entire
sensor probe area until the sensor and surrounding area is clear of
particles.
IMPORTANT: For excessive dust and/or foreign material buildup on or
around the temperature sensor, it is recommended the
instrument be returned to DICKEY-john Service.
Enlarged view
of sensor
probe
Summary of Contents for DNV25NB
Page 4: ...OPERATOR S MANUAL GAC 2500 Models UGMA INTL 11001 1505E 201305 IV ...
Page 6: ...OPERATOR S MANUAL GAC 2500 Models UGMA INTL 11001 1505E 201305 2 SAFETY NOTICES ...
Page 20: ...OPERATOR S MANUAL GAC 2500 Models UGMA INTL 11001 1505E 201305 16 NAVIGATION ...
Page 52: ...OPERATOR S MANUAL GAC 2500 Models UGMA INTL 11001 1505E 201305 48 SETUP ...
Page 72: ...OPERATOR S MANUAL GAC 2500 Models UGMA INTL 11001 1505E 201305 68 GRAIN CALIBRATIONS ...
Page 74: ...OPERATOR S MANUAL GAC 2500 Models UGMA INTL 11001 1505E 201305 70 DIAGNOSTICS ...
Page 86: ...OPERATOR S MANUAL GAC 2500 Models UGMA INTL 11001 1505E 201305 82 TROUBLESHOOTING ...