OPERATOR’S MANUAL
GAC® 2500 Models (UGMA & INTL)
11001-1505E-201305
SETUP / 35
RESULT SETUP
After performing a grain analysis, the unit can be set to automatically
advance back to the Sample ID screen from the Analysis Results screen or
manually return to this screen with a button press. The unit default is set to
manual return. Grain automatically dumps from the hopper in manual or
automatic advance mode.
To Display the Sample Return Setup screen:
1. At the System Setup screen, press the
More
button to display and
press the
Result Setup
button.
2. Select the desired feature to enable.
Automatic Return
Advances from the Analysis screen to the Sample ID screen within a
specified period of time (4 to 20 seconds) without pressing the
Initiate
button.
Manual Return
Advances to the Sample ID screen to perform another test by pressing the
Enter
button after each test.
Empty Drawer Requirement
A forced empty of the drawer requires a sample dump before proceeding to
the next analysis.
Bottomless Drawer in Use
Must be enabled when the bottomless drawer style is used with a
flow-through counter to prevent an empty drawer caution from occurring
when running tests.
– Grain can also be tested without the grain drawer secured in the
unit when Bottomless Drawer in Use is enabled; however, a Drawer
Ignored screen appears each time a test is performed and will only
proceed to perform test when the
Initiate
button is pressed.
Display Test Weight and Temperature
Displays the test weight and temperature on the Results screen for each
test. Both features are defaulted to enabled.
Summary of Contents for DNV25NB
Page 4: ...OPERATOR S MANUAL GAC 2500 Models UGMA INTL 11001 1505E 201305 IV ...
Page 6: ...OPERATOR S MANUAL GAC 2500 Models UGMA INTL 11001 1505E 201305 2 SAFETY NOTICES ...
Page 20: ...OPERATOR S MANUAL GAC 2500 Models UGMA INTL 11001 1505E 201305 16 NAVIGATION ...
Page 52: ...OPERATOR S MANUAL GAC 2500 Models UGMA INTL 11001 1505E 201305 48 SETUP ...
Page 72: ...OPERATOR S MANUAL GAC 2500 Models UGMA INTL 11001 1505E 201305 68 GRAIN CALIBRATIONS ...
Page 74: ...OPERATOR S MANUAL GAC 2500 Models UGMA INTL 11001 1505E 201305 70 DIAGNOSTICS ...
Page 86: ...OPERATOR S MANUAL GAC 2500 Models UGMA INTL 11001 1505E 201305 82 TROUBLESHOOTING ...