OPERATOR’S MANUAL
GAC® 2500 Models (UGMA & INTL)
11001-1505E-201305
SETUP / 33
SAMPLE SETUP
A Sample ID, Customer ID, User Name (Login) can be enabled so that a
grain sample is identified and saved with this information each time a test is
performed.
1. At the System Setup screen, press the the
More
button until the
Sample Setup
button appears.
2. Enable the check box of desired ID’s required when running samples.
3. Press the
Enter
button to accept.
Prompt for Sample ID
The Sample ID screen step can be hidden so that the screen does not
appear during analyzation. If a sample or customer ID is required, “Prompt
for Sample ID”, should be checked.
Prompt for Sample ID can be enabled even when a sample or customer ID
is not required so the screen appears during analyzation giving the option to
include either of these ID’s.
– Enable when a sample or customer ID are required entries before
analyzation. A sample ID will be associated with each grain test and
prints to ticket.
– Disable to eliminate the Sample ID screen before each analyzation
Sequence Sample ID
Automatic number sequencing to the next higher number will occur with
additional tests of the same grain. Sequence Sample ID default is enabled.
Customer ID Required
Requires a Customer ID to be associated with each grain test and prints to
ticket. Customer ID default is disabled.
User Login Required
A User Login requires a tester to login with a user name at the Main Menu
screen before grain can be analyzed. If a user login is required, the user
name entered at the User Login screen is recorded in the unit’s database
and prints on the print ticket. The
User
button only appears on the Main
Menu screen when enabled. User Login default is disabled.
Automatic Measure When Hopper is Full
Instrument automatically proceeds to measure a sample when the hopper is
full eliminating the need to press the
Initiate
button (green).
– Enable this feature to automate the measuring process. If a Sample
ID or Customer ID is required, analysis automatically starts at the
Sample ID screen. After the ID requirement is met, the
measurement automatically starts.
– Disable this feature to require a manual response at each analysis
screen.
Summary of Contents for DNV25NB
Page 4: ...OPERATOR S MANUAL GAC 2500 Models UGMA INTL 11001 1505E 201305 IV ...
Page 6: ...OPERATOR S MANUAL GAC 2500 Models UGMA INTL 11001 1505E 201305 2 SAFETY NOTICES ...
Page 20: ...OPERATOR S MANUAL GAC 2500 Models UGMA INTL 11001 1505E 201305 16 NAVIGATION ...
Page 52: ...OPERATOR S MANUAL GAC 2500 Models UGMA INTL 11001 1505E 201305 48 SETUP ...
Page 72: ...OPERATOR S MANUAL GAC 2500 Models UGMA INTL 11001 1505E 201305 68 GRAIN CALIBRATIONS ...
Page 74: ...OPERATOR S MANUAL GAC 2500 Models UGMA INTL 11001 1505E 201305 70 DIAGNOSTICS ...
Page 86: ...OPERATOR S MANUAL GAC 2500 Models UGMA INTL 11001 1505E 201305 82 TROUBLESHOOTING ...