OPERATOR’S MANUAL
GAC® 2500 Models (UGMA & INTL)
11001-1505E-201305
22
/
SETUP
CREATE PRODUCT
A new product can be created by copying a similar product’s parameters.
Creating a new product name and Issue ID prevents an existing product’s
parameters from being overwritten.
NOTE:
An external keyboard, barcode
scanner, or mouse can be used
to enter data into the unit.
To Create a Product:
1. At the View Product Setup screen, press the
Create
button.
2. At the Rename Product screen, the product name and NEW appears
as the default name.
3. Press the keypad icon to enter a different product name, if desired.
4. A new Issue ID is automatically defined for the new product. Press the
keypad icon to enter a different Issue ID, if desired.
5. Press the
Enter
button to save.
Figure 19
Creating a Product Screen
View Product Setup
View Product Setup
Product: Popcorn
Issue ID: 20101030
Calibration : UGMA
Range: 5 - 35%
Test Weight Range: 42-68 lbs/bu
Temperature Range: 25 - 125 ° F
Moisture Bias: -1.35
Moisture Slope: 1.00
Test Weight Bias: 0.00
Test Weight Slope: 1.00
File Name: Popcorn-20101030.CAL
Delete
Edit
Create
GAC2500-UGMA
Rename Product
Rename Product
New Product Name
Popcorn NEW
New Issue ID
20100116
Enter New Product Name
GAC2500-UGMA
Summary of Contents for DNV25NB
Page 4: ...OPERATOR S MANUAL GAC 2500 Models UGMA INTL 11001 1505E 201305 IV ...
Page 6: ...OPERATOR S MANUAL GAC 2500 Models UGMA INTL 11001 1505E 201305 2 SAFETY NOTICES ...
Page 20: ...OPERATOR S MANUAL GAC 2500 Models UGMA INTL 11001 1505E 201305 16 NAVIGATION ...
Page 52: ...OPERATOR S MANUAL GAC 2500 Models UGMA INTL 11001 1505E 201305 48 SETUP ...
Page 72: ...OPERATOR S MANUAL GAC 2500 Models UGMA INTL 11001 1505E 201305 68 GRAIN CALIBRATIONS ...
Page 74: ...OPERATOR S MANUAL GAC 2500 Models UGMA INTL 11001 1505E 201305 70 DIAGNOSTICS ...
Page 86: ...OPERATOR S MANUAL GAC 2500 Models UGMA INTL 11001 1505E 201305 82 TROUBLESHOOTING ...