Dynatron 900
+™
11
Intermittent Cycle
Longer Ramp Time
Harmonic Cycle
Static Cycle
A
DDITIONAL
T
RACTION
M
ODES
–
Q
UICK
S
ETUP
I
NSTRUCTIONS
S
TATIC
I
NSTRUCTIONS
1.
Prepare the patient for the traction treatment. Make sure all belts and harnesses are securely
attached.
2.
Plug in the remote stop cable and hand it to the patient.
3.
Turn on the Dynatron 900+ device.
4.
Set Treatment Time:
1-99 min.
5.
Set Hold Force Timer: “- -” sec.
6.
Set Hold Force:
2 - 91 kg. or 4 – 200 lbs.
7.
Adjust Speed
Fast
8.
Press Start
I
NTERMITTENT
I
NSTRUCTIONS
1. Prepare the patient for the traction treatment. Make sure all belts and harnesses are securely
attached.
2.
Plug in the remote stop cable and hand it to the patient.
3. Turn on the Dynatron 900+ device.
4. Set Treatment Time:
1-99 min.
5. Set Hold Force Timer: 1-99 sec.
6.
Set Rest Force Timer: 1-99 sec.
7.
Set Hold Force:
4 – 200 lbs. or 2 - 91 kg.
8.
Set Rest Force:
0 – 197 lbs. or 0 - 90 kg. Must be set
lower than Hold Force value.
9.
Adjust Speed
Fast:
Intermittent Cycle
Slower: Intermittent Cycle with
Longer Ramp Time
10.
Press START
H
ARMONIC
I
NSTRUCTIONS
1. Prepare the patient for the traction treatment.
Make sure all belts and harnesses are securely attached.
2.
Plug in the remote stop cable and hand it to the patient.
3. Turn on the Dynatron 900+ device.
4. Set Treatment Time:
(1-99 min.)
5. Set Hold Force Timer: (01 sec.)
6.
Set Rest Force Timer: (00 sec.)
7.
Set Hold Force:
(4 – 200 lbs. or 2 - 91 kg.)
8.
Set Rest Force:
(0 – 197 lbs. or 0 - 90 kg.) Must be set
lower than Hold Force value.
9.
Adjust Speed
Slow (minimum setting)
10. Press START