![Dynamica VELOCITY 18R Instruction Manual Download Page 4](http://html.mh-extra.com/html/dynamica/velocity-18r/velocity-18r_instruction-manual_2548263004.webp)
iii
Safety Reminder
WARNING:
This centrifuge is not explosion-proof. Never use explosive or flammable samples.
Do not install the centrifuge in or near places where inflammable gases are
generated or chemicals are stored.
Make sure to prepare necessary safety measures before using samples that are
toxic, radioactive or contaminated with pathogenic micro-organisms.
If the instrument, the rotor, and or accessories that has been contaminated by
solutions with toxic, radioactive or pathogenic materials, clean it according to the
decontamination procedure that you specified.
If the contaminated equipment requires services of DYNAMICA or authorized
agency of DYNAMICA, either a
t the customer’s site, DYNAMICA or at the agent
facilities, sterilize and decontaminate it in advance. Make sure to notify the service
representatives of the use of such materials.
Do not handle the power cord or turn on or off the POWER switch with wet hands
to avoid electrical shocks.
Users or any hazardous materials are recommended to keep 30cm away from the
centrifuge when it is operating.
While the rotor is rotating, never override the door lock.
Unauthorized repairs, disassembly, and other services maintenance applied to the
centrifuge except by our service personnel are strictly prohibited
。
When this product encounters disturbance of severe electrostatic discharge or
surge, it might have some degradation of performance like resetting or random errors
and so on. The operator should restart the equipment to recover the normal operation.
This is a class A product. In a domestic environment this product may cause radio
interference in which case the user may be required to take adequate measures.