![Dynamica VELOCITY 18R Instruction Manual Download Page 36](http://html.mh-extra.com/html/dynamica/velocity-18r/velocity-18r_instruction-manual_2548263036.webp)
30
13 List of applicable rotors and tubes
CAUTION:
To use the rotor properly please read the instruction manual carefully.
Do not run the centrifuge exceeding the allowable maximum speeds of the rotor,
buckets, and adapters. Some adapters, tubes and bottles have a lower speed than the
rotor.
13.1 Table of applicable rotors
Rotor type
Maximum
speed
Actual capacity
(ml×No. of tubes)
Tube/bottle
Maximum RCF
Part name
Size(
Φ×L)mm
FA18A
18,000rpm
27,070×g
10×8
10ml tube
Φ16×81
FA18B
18,000rpm
24,990×g
1.5/2.0×18
1.5/2ml tube
Φ10.5×41
FA15A
15,000rpm
21,500×g
1.5/2.0×24
1.5/2ml tube
Φ10.5×41
FA15B
15,000rpm
21,800×g
50×4
50 ml PP tube
Φ29×106
FA15C
15,000rpm
23,300×g
50×6
50ml PP tube
Φ29×106
FA15D
15,000rpm
20,900×g
80×4
80ml PP tube
Φ38×106
FA12A
12,000rpm
14,100×g
0.2×48
0.2ml tube
Φ6×24
FA10B
10,000rpm
11,400×g
50×6
50ml TC tube
Φ29×115
SW5
4,800rpm
4,170×g
250×4
250ml bottle
Φ61.6×135
SW5MP
4,800rpm
3,120×g
Plate×2
96 well microplate
96 Deep well plate
128×86×14
128×86×55