19
(5) Double press
button, and then the new parameters will be saved.
2) Recall programmed parameter.
(1) Turn on the power switch and set the rotor onto drive shaft.
(2) Press
button and the symbol
blinks, then the instrument enter a
programmed running mode.
(3) Press
or
button to choose your desired parameter. The parameters will
change with the serial number. The serial number is from 1 to 9 that is corresponding 1 to
9 group parameters respectively.
(4) Press
button to run the centrifuge, for details please refer to the section 7.1-5.
7.4 Pulse Operation
NOTE:
Under this mode, the acceleration and braking speed is designed to run at
maximum, regardless of the settings.
The centrifuge accelerates up to the setting speed.
If the
button is released during acceleration, the centrifuge will discontinue the
acceleration and starts the deceleration.
The
button works only when the rotor is stopped and when the door is locked.
1) Turn on the power switch and set the rotor onto the drive shaft.
2) Check the set speed and change it when necessary.
3) Press
button.
The rotor continues to accelerate while pressing this button, and when the centrifuge
reaches the set speed, it will continue to operate at the set speed.
4) Release
button.
The rotor starts the decelerating process, until it stops.
7.5 Pre-cooling operation
This function is designed to improve the ability to control the sample temperature.
Using this function, the operator can cool the rotor chamber and the rotor in advance.
1) Press
button, then the centrifuge displays pre-cooling screen.
The pre-cooling screen is shown as figure 7-5.