11 Troubleshooting
11.1 Common malfunction list
This centrifuge is designed with self-diagnosing function. For example, an error code of
the fault found will be displayed at the right bottom corner of the screen.
WARNING:
Do not open the door when the rotor is rotating.
Table 11-1 Error code
Symptom
Causes
Solution
Nothing appears on the
screen when the power
is turned on.
·Power socket blackout.
·The fuse is blown out.
·Remove the trouble and power
again.
·Replace the fuse.
Erro
r c
o
d
e
E
-x
x
a
p
p
e
a
re
d
o
n
t
h
e
ri
g
h
t
c
o
rn
e
r o
f
S
c
re
e
n
.
E-02
Door Open
·The door is open while the rotor
is rotating.
·The
button is pressed
while the door is open.
·Close the door immediately.
·Close the door, and then press the
button.
E-03
Rotor ID
The centrifuge can not identify
the rotor ID.
Reconfirm the ID code of the rotor
and make a correct selection.
E-04
Temp Ctrl
·Refrigeration ability is insufficient
·The air inlets on both sides of
the centrifuge are blocked.
·Too much dust is deposited on
the radiator.
·Something is wrong with the
refrigeration parts
·If the room temperature is high, lower
it.
·Dredge the air inlets on the
centrifuge.
·Clean the dust.
·Replace the damage part.
E-06
Over Set
The setting speed or g-value is
over the allowable range.
Check the settings and let it within the
allowable range.
E-08
No Rotor
Don
’t install the rotor
Install the rotor
E-09
Imbalance
·The imbalance is over the
allowable range.
·The machine is shocked during
operation.
·Balance the sample to ensure
imbalance is within the allowable
range.
·Don
’t push/move the instrument
when it is running.
E-10
~
86
Read Maintenance manual.
Contact
our
sales
or
service
representative. Inform them the alarm
code.
Error code E1 to E9 is mainly related to misoperation. User can continue using the
centrifuge after the malfunction is removed.