BlueWave® MX-Series Multichannel Controller User Guide
Irradiating Time (DB) (1-4):
The irradiating time for each channel is displayed
in the second column under the heading T(s). When in timed mode, time is
entered by touching the T(s) button, which in turn brings up a numerical
keypad for entry. The entry range is 0.1
–
999 seconds. From 0.1
–
30 seconds
the adjustment is made in 0.1 second increments and from 30
–
999 seconds
the adjustment is made in 1 second increments.
The button will display a colored highlight in the following situations:
Yellow highlight emitter time is waiting for delay time to expire.
Green highlight emitter is ON.
Red highlight Emitter Alarm.
No Highlight emitter ready.
Emitter Power (DB) (1-4):
The cure power level is displayed as a percentage
in the third column, under the heading P(%). Pressing this display field will
advance the unit to the power level screen. Power may be set from 1-100% in
1% increments.
Note: Always use the lowest power setting to achieve cure to extend emitter
life.
Emitter Delay (DB) (1-4):
The cure delay is displayed in the fourth column
under the heading D(s). This is a delay between the emitter trigger being
activated and the emitter irradiation. Pressing this display field advances to the
delay time screen.
Emitter Trigger (DB)(1-4):
The emitter trigger button is located in the fifth
column under the heading Tr. It displays the trigger number used by each
emitter to start the curing process. When this button is pressed, the trigger
number will increment up. By setting common trigger number, multiple
emitters can be triggered simultaneously by a single trigger input.