15
Dymax BlueWave® LED VisiCure® User Guide
This output can be used to verify step or ramped changes in the intensity for customized profiles, or to
alert the remote controller when the system has reached maximum intensity in a closed-loop system.
Pin 9
–
Over Temp Out:
A sensor constantly monitors the temperature of the LED in the system. If, for
any reason, the reading rises above a predetermined limit, the control PCB will shut off current to the LED
to prevent damage. When the measured temperature drops to a lower threshold, the PCB will re-enable
current to the LED.
NOTE: If this fault occurs, check for proper airflow into and out of the unit. Also verify that the air filter is
clean.
Figure 9. 9-Pin D Connector I/O Signals
Cleaning and Maintenance
The BlueWave LED VisiCure was designed to operate with minimum maintenance. Follow the schedule
below to assure top unit performance.
Lightguide
Clean the lightguide ends monthly or as required. The ends of the lightguide should be kept clean to
transmit as much light as possible. Cured adhesive can be carefully removed from liquid-filled lightguide
ends with a razor blade. Avoid sharp bends with the lightguide since this reduces light output and
permanently damages the guide.