EN
2012/02
23
25
26
•Activatethesuctionunitbyremovingasuc-
tionhosefromthehosemanifoldorbyoper-
atingtherinsingphaseofthespittoon(spit-
toonvalve,"WetSuctionSystem")
•Carryoutafunctioncheckandtestallcon-
nectionsforleaksetc
Check AC suction units for correct
direction of rotation.
•Switchoffthesuctionunit,eitherbyreplacing
asuctionhoseintothehosemanifoldorby
switchingthespittoonrinsingoff
VS-Suctionunitsrunforapprox30
secondsafterswitchingoff
Simultaneoussignalstostartfromboth
thesuctionunitandthecompressors,
causethecompressorstobeginaftera
slightdelayafterthesuctionunitstarts
up
•Hangallpanels(26)ofthePTS200intoplace
Danger of overheating
The PTS must be closed once all
panels are in position. This serves
to provide sufficient ventilation in
the PTS.
a
26