background image

7

when the Meter is turned off.

Button

Function

SELECT Disables automatic power-off function. Display shows PoFF until the 

button is released.

ENTER

Disable the “beep” function.

• 

The Meter has automatic power-off function (sleep mode), which conserves 

battery power if there is no function change or button press for ten minutes.

• 

The Meter comes out of Sleep mode when a key is pressed or when the rotary 

switch is changed.

• 

To disable Sleep mode, hold down the SELECT button while turning the Meter on. 

• 

Sleep mode is always disabled in the ‘Insulation Test’ mode, or if the auto power-

off feature has been disabled by pressing SELECT when turning the Meter on.

AUTOMATIC POWER-OFF (SLEEP MODE)

HOLD FUNCTION

• 

Press the HOLD button to freeze the displayed value. Press again to release the 

display.

RELATIVE MEASUREMENT

• 

Show the difference between the actual value and the relative base.

• 

Press the REL button to enter relative measurement and the Meter will record the 

initial value when pressing the key.

• 

Displayed value = Actual value - Initial value.

• 

Press the REL button again to exit relative measurement.

• 

Press the SELECT button to view Initial value, when the Initial value is invalid the 

display will read ----.

 

• 

In Insulation test mode, press the TEST button to perform an insulation test until 

the button is released. When the button is released, the screen will display the 

“Hold” sign.

• 

Press the LOCK button, then the screen will display the “Lock” sign.

• 

Press the TEST button and the Meter will perform an insulation test until you press 

the TEST button again.

• 

Lock function is invalid when the timer function is selected.

INSULATION TEST LOCK

STORING TEST DATA

• 

Press the MEMO button so that the meter enters HOLD status automatically.

• 

The LCD screen will display “MEMO” and the memory code on the secondary 

display.

• 

Press    /    button to change the code, press enter to save the data in the preferred 

memory code and the screen will display “SAVE”.

Summary of Contents for D03131

Page 1: ...1 3 DIGITAL MULTIMETER MODEL D03131...

Page 2: ...ons 7 Automatic Power Off 7 Hold Function 7 Relative Measurement 7 Insulation Test Lock 7 Storing Test Data 8 Reading Test Data 8 Delete Data 9 Compare Function 9 Timer Function 9 Max Min Avg 9 DAR PI...

Page 3: ...own voltage Do not apply more than the rated voltage as marked on the Meter between the terminals or between any terminal and earth ground Use caution with voltages above 30V AC RMS 42V AC peak or 60...

Page 4: ...ay COMP Compare function is selected PASS When the measured value is between the upper limited and lower limited value REL Relative function is selected DAR In insulation test mode display the DAR val...

Page 5: ...ion Replace the battery bat Appears on auxiliary display indicating that the battery is too low to perform the insulation test P r E S Preset value POFF Auto power off function is disabled LIVE In ins...

Page 6: ...position 9 Activate the compare function 10 Select one test output voltage for insulation test 11 Activate the Insulation test ROTARY SWITCH Position Function OFF Turn off the Meter power V DC Voltage...

Page 7: ...relative base Press the REL button to enter relative measurement and the Meter will record the initial value when pressing the key Displayed value Actual value Initial value Press the REL button agai...

Page 8: ...he READ button to display the data in memory Press to change the code and read the data accordingly DELETE DATA In READ mode press ENTER and the meter will display dEL n Press ENTER again to delete th...

Page 9: ...ER The lock function will be invalid and the Meter will start the Insulation test when pressing the MEASURE button When the time is due the test is stopped In Timer mode the current voltage and time i...

Page 10: ...nuity test features a beep that sounds as long as the circuit is complete The beep sounds when a short 3 INSULATION TEST Warning When testing voltage output do not rotate the rotary switch to prevent...

Page 11: ...tion of test Insulation Maximum Capacitive Load Operable with up to 1 F load Storage Temperature 40 C to 60 C Operating Temperature 0 C to 40 C Storage Altitude 12000m Operating Altitude 2000m 1000V C...

Page 12: ...SAL FOR CONSUMERS OF ELECTRICAL ELECTRONIC EQUIPMENT When this product has reached the end of its life it must be treated as Waste Electrical Electronics Equipment WEEE Any WEEE marked products must n...

Reviews: