MEASURING RESISTANCE

WARNING

• 

Beware of electrocution.

• 

When measuring in-circuit resistance, make sure that the power of the circuit 

under test has been turned off and that all capacitors have been fully discharged.

• 

Plug the black test lead into the 

COM

 jack and the red test 

lead into the 

INPUT

 jack.

• 

Set the rotary selector to the      range position to ensure that 

the meter goes into ‘

Ω

’ range.

• 

Connect the test leads to the ends of the resistor or circuit 

for measurement.

• 

Take the reading on the LCD.

Note: When the input is open, the “OL” will appear on the LCD to 

indicate that the range has been exceeded.

• 

For measuring resistance above 1MΩ, it may take a few 

seconds to get a steady reading. This is normal for high 

resistance reading.

 

TESTING DIODE

• 

Plug the black test lead into the 

COM

 jack and the red test lead into the 

INPUT

 

jack.

• 

Set the rotary selector to the      range position.

• 

Press the “

SEL

” button to switch to      test. 

• 

Connect the red test lead to the anode and the black test lead to the cathode of 

the diode for testing.

• 

Take the reading on the LCD.

Note: The meter will show the approximate forward voltage drop of the diode.

When the test leads have been reversed or opened, “OL” will appear on the LCD.

 

TESTING CONTINUITY

WARNING

• 

Beware of electrocution.

• 

Make sure that the power of the circuit has been turned 

off and the capacitors have been fully discharged before 

testing the continuity of a circuit.

• 

Take the reading on the LCD.

Note: If the duty cycle is less than 10%, the symbol “UL” will be displayed on the LCD. 

If the duty cycle is more than 99.9%, the symbol “OL” will be displayed on the LCD.

• 

The input signal frequency range is 10 - 10kHz. It is possible to test the duty cycle 

of a frequency signal higher than 10kHz, but the tolerance of the test result can’t 

be ensured.

• 

“    ” means the maximum input voltage is 750V rms AC.

• 

Plug the black test lead into the 

COM

 jack and the red test 

lead into the 

INPUT

 jack.

• 

Set the rotary selector to the      range position.

• 

Press the “

SEL

” button to switch to      continuity test.

• 

Connect the test leads to the two ends of the circuit for 

 

Summary of Contents for D03128

Page 1: ...1 CLAMP METER MODEL D03128...

Page 2: ...Frequency or Duty Cycle 9 Switching Maximum or Minimum Value 10 Switching Functions 10 Backlight and Clamp Lighting Bulb 10 Auto Power Off 10 Preparing for Measurement 11 Measuring AC Current 11 Measu...

Page 3: ...Do not use the meter around explosive gas or vapour When using the test leads keep your fingers behind the finger guards Remove test leads from the meter before opening the meter case or battery door...

Page 4: ...12 Reading Hold Backlight Button HOLD B L 3 Panel 13 Rotary Selector 4 Trigger 14 OFF Power Switch 5 Function Switch Button SEL 15 Symbol 6 MAX MIN Switch Button MAX MIN 16 Symbol 7 Relative Switch Bu...

Page 5: ...INPUT JACKS Button Switch Jack Function HOLD B L button Hold the reading or control the backlight SEL button For switching among measuring functions REL button Relative value measurement Hz button Sw...

Page 6: ...rate 3 times sec Bar graph 30 times sec Unit indication Function and unit Auto power off time 15 minutes Operating power 1 5V x 3 AAA batteries Battery low indication on LCD Temperature factor 0 1 x...

Page 7: ...ue Frequency by a range from current clamp Range Resolution Accuracy 10Hz 0 01Hz 1 5 of rdg 5 digits 1kHz 0 001kHz 1kHz 0 001kHz Measurement range 10 1kHz Input current range 4A rms AC higher input cu...

Page 8: ...of rdg 3 digits 4k 0 001k 40k 0 01k 400k 0 1k 4M 0 001Mk 40M 0 1M 1 2 of rdg 3 digits Open circuit voltage 0 4V Overload protection 250V DC or rms AC Diode Range Resolution Function 0 001V Displaying...

Page 9: ...mode OL shows when the input value is larger than the allowed value of the measurement mode Press the key again and the relative measurement function will be cancelled Disable to enter REL mode when...

Page 10: ...set at about 15 seconds meaning the light will turn off automatically after about 15 seconds frequent use of the backlight will shorten the life of the batteries When the battery voltage is 3 7V the...

Page 11: ...400A rms AC MEASURING DC CURRENT WARNING Beware of electrocution Ensure that the test leads are disconnected from the meter before making current clamp measurements Set the rotary selector to the 40A...

Page 12: ...e that the test leads are disconnected from the meter before making current clamp measurements Set the rotary selector to the A range A or A position Press the trigger to open the jaw and fully enclos...

Page 13: ...uency test range is 10Hz 10kHz It is possible to test the frequency which is higher than 10kHz but the tolerance of the test result can t be ensured means the maximum input voltage is 750V rms AC WARN...

Page 14: ...V range position Press the Hz button to switch to DUTY measurement Connect the test leads to the two ends of the source or load for measurement Take the reading on the LCD Note If the duty cyle is les...

Page 15: ...ed test lead to the anode and the black test lead to the cathode of the diode for testing Take the reading on the LCD Note The meter will show the approximate forward voltage drop of the diode When th...

Page 16: ...es If the low battery symbol appears it means that the batteries should be replaced Loosen the fixing screw of the battery cover and remove it Replace the exhausted batteries with new ones Put the bat...

Reviews: