7
DC Voltage
Range Resolution
Accuracy
400mV
0.1mV
± (1.0% of rdg + 2 digits)
4V
0.001V
± (0.7% of rdg + 2 digits)
40V
0.01V
400V
0.1V
600V
1V
± (0.8% of rdg + 2 digits)
•
Input impedance: 10MΩ.
• Max. input voltage: 1000V DC.
• At small voltage range, unsteady
readings will appear before the test
leads contact the circuit. This is
normal because the meter is highly
sensitive. When the test leads
contact the circuit, the true reading
will be shown.
AC Voltage
Range Resolution
Accuracy
4V
0.001V
± (0.8% of rdg + 3 digits)
40V
0.01V
400V
0.1V
750V
1V
± (1.0% of rdg + 4 digits)
•
Input impedance: 10MΩ.
• Max. input voltage: 750V rms AC.
• Frequency range: 40 to 400Hz.
• Response: Average value.
Frequency
by a range (from current clamp)
Range Resolution
Accuracy
10Hz
0.01Hz
± (1.5% of rdg + 5
digits)
1kHz
0.001kHz
>1kHz 0.001kHz
• Measurement range: 10 ~ 1kHz.
•
Input current range: ≥ 4A rms AC
(higher input current at higher
frequency).
• Max. input current: 400A rms AC.
Frequency
by ACV range
Range Resolution
Accuracy
10Hz
0.01Hz
± (1.5% of rdg + 5
digits)
1kHz
0.001kHz
10kHz
0.01kHz
>10kHz 0.01kHz
Take it only as reference
• Measurement range: 10 ~ 10kHz.
•
Input voltage range: ≥ 0.6V rms
AC (higher input voltage at higher
frequency).
•
Input impedance: 10MΩ.
• Max. input voltage: 750V rms AC.
Frequency
by Hz/Duty Cycle range
Range
Resolution
Accuracy
9.999Hz
0.001Hz
± (0.5% of rdg + 3
digits)
99.99Hz
0.01Hz
999.9Hz
0.1Hz
9.999KHz 0.001kHz
99.99KHz 0.01kHz
999.9KHz 0.1KHz
9.999MHz 0.001Mhz
• Overload protection: 250V DC or
250V AC rms.
• Input voltage range: 200mV-10V
AC rms.