D U N I W A Y S T O C K R O O M C O R P .
10 of 30
II
Installation
A. Unpack the Controller
Carefully unpack the Model 960 Dual - Hybrid Vacuum Gauge Controller. The shipment includes
these components:
• controller unit
• power cord
• mounting clips
• D-sub 15 accessory connector
• this instruction manual
If your controller does not have all of these items, call Duniway Stockroom. If anything appears
to have been damaged in shipment, contact the shipper.
B. Mount the Controller
You can rest the controller unit on a bench, table top, or shelf, or you can mount it in a rack or
cabinet. The controller unit is housed in a standard 1/8 DIN box. If you are mounting the unit in
a panel, the cutout dimensions are 1.78 inch by 3.60 inch (45.2 mm by 91.4 mm), see fig. 3, page
9. One mounting clip attaches to each of the sides of the controller unit. To attach the clip, slide
the beveled surfaces of the clip under the cutout on the side of the box and push the clip toward
the back of the unit.
Be sure to leave enough clearance at the back of the controller unit for easy access to cable con-
nections and enough space around the vents of the unit to allow heated air to escape.
C. Select the CCG
The Model 960 controller is designed to work with the Duniway CCG-525 and Varian Model 525
Cold Cathode gauges.
D. Select the CVT Tube
The Model 960 controller is designed to work with the CVT-275 or the
Brooks/Helix/Granville-Phillips type 275 Convectron
®
sensor tubes, both of which are supplied
by Duniway Stockroom Corp.
TELEPHONE: 650-969-8811 TOLL-FREE (U.S. only): 800-446-8811 FAX: 650-965-0764 EMAIL: [email protected]
www.duniway.com