D U N I W A Y S T O C K R O O M C O R P .
7 of 30
5. Full Scale
CCG:
1.0 x 10
-2
torr
CVT
995 Torr (mBar)
6. Calibration Adjustments:
CCG:
CCG Calibrate
UNITS
CVT:
GAUGE ZERO adjust
UNITS
CONV. ATM adjust
7. Vacuum Gauge
CCG:
one Duniway CCG-525 or Varian 525 Cold Cathode gauge
CVT
one CVT-275 or G-P 275 CONVECTRON
®
.
8. Operating Temperature Range
+2 to +50 deg. Celsius
9. Process Control Set Points
two, with independent High and Low set points for each relay, for flexible control of hysteresis,
assignable to either sensor
10. Process Control Relays
two relays; contacts rated at 2 amp/240
VAC
, 300
VDC
11. Nonvolatile Memory for all user specified parameters
12. Analog Output
logarithmic, 0.5 volts/decade; 0.1 mtorr = 4.00 volts
13. Output Voltage
CCG
-2 kv
TELEPHONE: 650-969-8811 TOLL-FREE (U.S. only): 800-446-8811 FAX: 650-965-0764 EMAIL: [email protected]