8
1. Detection interval:
no more than 24 hours and no less than 60 seconds. Note: When performing scheduled
detection, selecting the corresponding channel under the status display bar is needed before it takes effect.
2. Threshold of all zones:
alarm threshold of impedance change for all zones (20~99% is light load, above 100%
is overload, 100% is normal). You can also set the impedance change threshold for each partition separately under
the Status display bar. Tap the number in the light load or overload and press ENTER when the settings are
complete to save the settings. Note: threshold set for all zones will override threshold for each zone.
3. Backlight time: click
“
”, five options– 1 minute, 10 minutes, 30 minutes, 60 minutes, always on.
4. Language:
click “
” to change language.
5. Tap sound:
to set tap sound.
6. Alarm sound:
to set whether to give out alarm sound when fault occurs in zones.
7. All zone calibration:
open this button to calibrate all zones. It will close automatically.
Note
: all zone calibration is needed at first time use or after repair and maintenance.
8. System time system:
click “
” to enter interface as Figure 3-3,.
9. Help:
click “
” to enter the interface as Figure 3-2.
Figure 3-2