![DS Maref DVT-2600 User Manual Download Page 36](http://html.mh-extra.com/html/ds-maref/dvt-2600/dvt-2600_user-manual_2533352036.webp)
*Normal range: 9.5~ 13.0v
Main board program
version
Sub board program
version
*Normal range: 14.5~15.5v
Test number
Test number
Test number
When you press ‘START’,
you can turn on or turn out
the ‘Lymph’ mode.
Test number
Test number
Test number
Test number
*Normal range: 9.5~ 13.0v
Main board program
version
Sub board program
version
*Normal range: 14.5~15.5v
Test number
Test number
Test number
When you press ‘START’,
you can turn on or turn out
the ‘Lymph’ mode.
Test number
Test number
Test number
Test number
Maintenance and safety instructions
7
7-1. Test methods and calibration
How to use
When you press 'LEG' and 'MODE' buttons for
2 seconds at the same time, test method mode
including 10 test programs starts. Whenever
you press the 'mode' button, the test number is
changed from 1 to 10 and when you press the
'START', you can change the selected value or do
the test. If you want to finish test method, please
press 'START' for 3 seconds.
The devices maintenance and correction should be followed by
certified technician's instruction.
Check the program version
To remove the lymph mode
Check voltage for charging
battery
Check battery voltage
1
2
3
4
36