For sleeves with 3 chambers such as the leg/calf/boot sleeve; if 40mmHg is set, pressure
Will be delivered as 45 / 40 / 35 mmHg to the three chambers.
4-5-2. Foot Pressure Button
This button sets the pressure at which the device will be operated.
The pressure setting cannot be changed during operating.
Basic pressure is set as 120mmHg. The pressure setting can be changed by 10mmHg
Increments by pressing the button(ex. 120 130 140 120mmHg).
The device automatically selects the sleeve when the START/STOP button is selected.
4-5-3 Interval Button
This button sets the interval time. The time value which the user sets will be saved. It can
be changed to 48 - 60 - 28 - 48 sec each time the interval button is selected. After initial
inflation of pressure to each chamber the device will then restart therapy with the interval
time selected. This operation will be repeated and the setting cannot be changed during
operation.