Drucker Diagnostics – Customer Service: +1-814-692-7661 – [email protected] | Page 15
C
ALCULATING THE
G-F
ORCE
The I.F.U.s of tube manufacturers recommend cycles at a minimum G-Force, which can be calculated if you know
the RPM and the radius. Use the formula below or go to
www.druckerdiagnostics.com/g-force-calculator/
In Centimeters:
RCF or G-force = 0.00001118 x
Rotor Radius (cm) x (RPM)
2
In Inches:
RCF or G-force = 0.0000284 x
Rotor Radius (in) x (RPM)
2
Radius
2.8 in (7.1 cm)
R
EPLACEMENT
P
ARTS
Part No.
Description
7724177
Foot, rubber
02-002-1-0044
Lid Tray Assembly
420545
12 Place Rotor Assembly
420526
6 Place, 100mm Rotor Assembly
03-1-0006-0008
Motor Assembly
02-006-0-0021
PC Board
7760006
Power cord
03-1-0005-0193
Power Supply, 350W
02-002-1-0041
Lid Assembly
7724071
Hinge, friction
02-002-1-0057
Seal, lid gasket
03-0-0003-0332
Open/Close Label
02002-1-0075
Lid LED Assembly, Yellow
03-0-0003-0577
Sero 12 Front Panel Label
03-1-0001-0090
Button Cover No Emboss
03-1-0001-0089
Button Cover Embossed
03-1-0002-0093
Button Spacer
031-0002-0095
Button Protector
00-100-100-005
Replacement Grommet & Bushing Kit