Appendix 1: Troubleshooting

1) In the period of using Smart series microscope, if there is any trouble occurs, please referring to

the following sheet listed some common troubleshooting resolve them.

Trouble

Causation

Remedy

Switch on but bulb dark

Plug is unreliable

Plug in again

Bulb is broken

Change bulb

Fuse is broken

Change fuse

Bulb is flickering or

brightness is unsteady

Bulb is unstable

Insert it again

Bulb is broken

Replacing bulb

Brightness of view field

isn’t enough or is uneven

Bulb specification doesn’t

meet the requirement

Replacing bulb

Brightness isn’t adjusted

correctly

Adjust rotation

potentiometer

Objective isn’t in correct

position

Make the objective in correct

position

The size of iris aperture is too

small

Adjust the size of iris aperture

Brightness of view field

isn’t enough or is uneven

Lens (objective, eyepiece,

condenser, light collector) has

dust

Clean it

Position of condenser is too low Higher condenser

Image isn’t clear

(contrast or definition

isn’t enough)

Cover glass of specimen

doesn’t meet the requirement

Use required thickness cover

glass (0.17mm)

Cover glass of specimen isn’t in

up direction

Place specimen correctly

Surface of objective lens

Is dirty (especially it is easy for

the front lens of 40

×

objective

to dip in immersion oil)

Clean it

Immersion oil isn’t used for

100

×

objective (oil)

Use immersion oil

Immersion oil doesn’t meet the

requirement

Use immersion oil

supplied by us

There is bubble in immersion oil Clear the bubble way
Size of iris aperture isn’t

proper

Adjust the size of iris

aperture

Position of condenser is too low Readjust the position of

condenser

One side of image is

dark or image is

moving as focusing

Objective isn’t in correct

position

Make the objective in

correct position

Specimen isn’t placed correctly Place specimen levelly on stage

and clip it with clamp

Objective touches

specimen as changing low

times objective to high

times objective

Cover glass of specimen isn’t in

up direction

Place specimen correctly

Cover glass doesn’t meet the

requirement

Use required thickness

cover glass (0.17mm)

Image observed by two

eyes aren’t in

superposition entirely.

Interpupilary distance isn’t

adjusted correctly.

Adjust interpupilary

distance according to two eyes

It is easy for eyes to be

tired during observing

Diopter isn’t adjusted correctly Readjust diopter

2) If there is any trouble occurs in digital system , please referring to the following sheet listed some
common troubleshooting resolve them.

Summary of Contents for SMART-e Series

Page 1: ...operation Drawell International Technology Limited Shanghai Drawell Scientific Instrument Co Ltd Add Suite 1506 Lane581 XiuChuan Rd Greeland Max Mall PuDong Shanghai China Tel 0086 21 50966080 Fax 008...

Page 2: ...SMART e Series Digital Microscope Operation Instruction To ensure the safety and obtain satisfactory performance please study this operation instruction thoroughly before your operation...

Page 3: ...And Carry 4 4 Operation And Use 5 5 Image System Operation 8 6 Maintenance 9 Appendix1 Troubleshooting 11 Appendix2 Digital System Technical Parameter 12 Appendix3 Standard Outfits 13 Please read the...

Page 4: ...Screw or knurled thumb screw H Mechanical Stage Moving vertically and horizontally Knob J Right Coarse Fine Focusing Knobs K Brightness control knob F Body handle L Power switch M Condenser Focusing K...

Page 5: ...nal Fine division 2 m Condenser up down range 26mm 7 Condenser Abbe condenser N A 1 25 Adjustable aperture Aperture center can be adjustable 8 Illumination Non spherical illumination System Field Diap...

Page 6: ...n condenser screw raising condenser installing ring to the top Electrical Equipment Checking Plug power line in power input turn on power switch L observe brightness of light collector then continue t...

Page 7: ...comfortable sensation of the coarse and fine focusing knob prevent moving downward Clockwise can relax instead can lock 5 Adjust condenser aperture diaphragm Adjust condenser focusing knob M and aper...

Page 8: ...Rotate condenser focusing knob until in the middle of the field 3 Put 40 Objective in optical path rotate E will make field image slightly bigger than field Please readjust if field image isn t in mid...

Page 9: ...to pull out power wire from charging power socket in order to prevent the backflow of discharge battery electric current 3 Charging machine please replace 3 batterys in same size when battery can t c...

Page 10: ...o browse image collector function clear or basic clear image should also be on the screen if isn t satisfied only slightly rotating J N micro focusing hand wheel until get satisfactory imaging results...

Page 11: ...and in carrying Don t drag it on the surface of worktable to avoid damage to microscope and worktable 6 3 Replacement of bulb 6 3 1 Replacement of halogen Please use specified bulbs If use other ligh...

Page 12: ...t cut off power of microscope and pull out the power plug 2 Take off fuse base in the back of base take out old fuse 3 Replace a new fuse with same size then plug fuse base in 6 5 When not in use 1 Pl...

Page 13: ...ace specimen correctly Surface of objective lens Is dirty especially it is easy for the front lens of 40 objective to dip in immersion oil Clean it Immersion oil isn t used for 100 objective oil Use i...

Page 14: ...Adjusting diopter adjusting ring Software calibration error Incorrect operation procedure Please carefully reference manual procedure Calibration select error Among the image and to gauge calibration...

Page 15: ...0 Rotatable 360 Interpupilary Distance 50 75mm Nosepiece Quadplex Mechanical Stage Stage Size 145X140mm Travel 76mm 52mm Scale 0 1mm Two Slide Holder Condenser Abbe N A 1 25 Iris Diaphragm Illuminatio...

Page 16: ...ology Limited Shanghai Drawell Scientific Instrument Co Ltd Add Suite 1506 Lane581 XiuChuan Rd Greeland Max Mall PuDong Shanghai China Tel 0086 21 50966080 Fax 0086 21 50966080 Web www drawell com cn...

Reviews: