4.

Operation And Use

4.1 Bright field operation process instruction

Cover slip: using cover slip of thickness 0.17mm can get ideal working environment for objective and
reach design performance.

Slide glass: using slide glass of thickness 1.2mm(0.9

1.4mm) for better image.

Avoid reverse rotating left and right coarse & fine focusing knob together, otherwise it will damage
focusing unit.

Don’t switch objective directly when changing objective, should rotate the tooth ripple on nosepiece to
put objective in optical path.

Interpupillary distance is different from person to person, so readjustment each time.

Phenomena of automatic mechanical stage moving downward may occur after long-term use . Tension
adjustment ring can adjust firmness and comfortable sensation of the coarse and fine focusing knob,
prevent moving downward. Clockwise, can relax, instead, can lock.

(5)Adjust condenser aperture

diaphragm

Adjust condenser

⑤(

focusing knob M

and aperture (Aperture diaphragm lever D).

Observing

(1) Switch on the power,

adjust brightness.

Turn on power switch L, adjust brightness knob K.

(2)place specimen .

Put slices on stage

, (

open C slices handle, place

slices)and adjust H stage moving handle, put specimen
in the optical path.

(3)Put 10

×

objective into optical

path and focus.

Turn Nosepiece B, put 10

×

objective into optical path.

Turn Coarse & Fine Focusing Knobs J and N to focus.

(4)Adjust interpupillary distance

and diopter

Adjust intergration head

to make two

fields into one; adjust diopter adjustment ring
A to make diopter for two eyes same.

(6) Adjust and center field

diaphragm

Adjust field diaphragm (field diaphragm ring
E ) (model with field diaphragm).

Put the objective into

light path, and focus.

Summary of Contents for SMART-e Series

Page 1: ...operation Drawell International Technology Limited Shanghai Drawell Scientific Instrument Co Ltd Add Suite 1506 Lane581 XiuChuan Rd Greeland Max Mall PuDong Shanghai China Tel 0086 21 50966080 Fax 008...

Page 2: ...SMART e Series Digital Microscope Operation Instruction To ensure the safety and obtain satisfactory performance please study this operation instruction thoroughly before your operation...

Page 3: ...And Carry 4 4 Operation And Use 5 5 Image System Operation 8 6 Maintenance 9 Appendix1 Troubleshooting 11 Appendix2 Digital System Technical Parameter 12 Appendix3 Standard Outfits 13 Please read the...

Page 4: ...Screw or knurled thumb screw H Mechanical Stage Moving vertically and horizontally Knob J Right Coarse Fine Focusing Knobs K Brightness control knob F Body handle L Power switch M Condenser Focusing K...

Page 5: ...nal Fine division 2 m Condenser up down range 26mm 7 Condenser Abbe condenser N A 1 25 Adjustable aperture Aperture center can be adjustable 8 Illumination Non spherical illumination System Field Diap...

Page 6: ...n condenser screw raising condenser installing ring to the top Electrical Equipment Checking Plug power line in power input turn on power switch L observe brightness of light collector then continue t...

Page 7: ...comfortable sensation of the coarse and fine focusing knob prevent moving downward Clockwise can relax instead can lock 5 Adjust condenser aperture diaphragm Adjust condenser focusing knob M and aper...

Page 8: ...Rotate condenser focusing knob until in the middle of the field 3 Put 40 Objective in optical path rotate E will make field image slightly bigger than field Please readjust if field image isn t in mid...

Page 9: ...to pull out power wire from charging power socket in order to prevent the backflow of discharge battery electric current 3 Charging machine please replace 3 batterys in same size when battery can t c...

Page 10: ...o browse image collector function clear or basic clear image should also be on the screen if isn t satisfied only slightly rotating J N micro focusing hand wheel until get satisfactory imaging results...

Page 11: ...and in carrying Don t drag it on the surface of worktable to avoid damage to microscope and worktable 6 3 Replacement of bulb 6 3 1 Replacement of halogen Please use specified bulbs If use other ligh...

Page 12: ...t cut off power of microscope and pull out the power plug 2 Take off fuse base in the back of base take out old fuse 3 Replace a new fuse with same size then plug fuse base in 6 5 When not in use 1 Pl...

Page 13: ...ace specimen correctly Surface of objective lens Is dirty especially it is easy for the front lens of 40 objective to dip in immersion oil Clean it Immersion oil isn t used for 100 objective oil Use i...

Page 14: ...Adjusting diopter adjusting ring Software calibration error Incorrect operation procedure Please carefully reference manual procedure Calibration select error Among the image and to gauge calibration...

Page 15: ...0 Rotatable 360 Interpupilary Distance 50 75mm Nosepiece Quadplex Mechanical Stage Stage Size 145X140mm Travel 76mm 52mm Scale 0 1mm Two Slide Holder Condenser Abbe N A 1 25 Iris Diaphragm Illuminatio...