10
6.2 How to Operate
6.2.1 Rotor Installation
The drive spindle in the chamber and the inside cones of rotor should be first
wiped with a piece of soft and clean cloth, then place the rotor with two hands
to the cone side of the drive spindle in the centrifuge chamber, Washer should
be placed before tightening the screw nuts.
6.2.2 Tubes Installation
Place the bracket into the slot of rotor; make sure that the bracket is turning
smoothly, tube sets (adaptors) are placed into holes of brackets, the solution to
be separated into the tubes. Visually check if the solution in the tubes is
roughly equal before putting them into the tube sets.
6.2.3 Preset
Fig.1 Control Panel
1) Function Declaration of Display Window
①
Rotor number Display Window:
press "
SET
" ,when power indicator (red)light
is on, the window of rotor number is flashing, turn the "
ENTER"
key to
confirm the number.
②
Speed Dispaly Window:
It can display the running speed and RCF.
Note: the set speed must not exceed the max. speed limit.
③
Time Display Window:
The window can display the time and the Erro, Detailed
reference to the following sheet 2.
④
Set:
This button can set rotor number, speed and time.
2) Set button :
can set rotor number, speed and time.