background image

5.2. The Max weighing is the showing largest value after turn on add 9d

(e=10d,”d” is minimum reading). If overloading,it will show “------”and you

must take away the goods at once, or the balance (load cell,especially) will

be broken.

6. Tare

Press

, the balance pan can be tared if the blank spot disappears.

But if there is the black spot, the tare function is no effect.

7. Backlight

Press

to turn on/off the backlight.

8. Zero-Tracking and Auto Tare

8.1. Turn off the balance, one hand press and hold

at the same time,

another hand turn on the balance again, “-Zero-” will show, then loose

11. Maintenance

Failure

Cause

Remedy

No light

·Scale is not connected with

power

·Power switch is off

·Adapter is broken

·PCB is broken

·To plug in the power

·To press

·To change new adapter

·To change new PCB

Upper “----”

only

·over-load

·load cell wire loose

·To decrease load

·To re-connect the wire

S-CAL

·load cell wire loose

·No connect load cell wire

·To connect load cell wire

Value

unstable

·Big air

·Working stage is unstable

·Big

change

of

room

temperature

·Load cell is broken

·To shut door or window

·To put scale on stable

stage

·To control temperature

·To check load cell

Value

inaccuracy

or big error

·No clear zero before weight

·No calibrate or calibration

weight is inaccuracy

·Wrong voltage

·Big error in four corners of

load cell

·Load cell is broken

·To press

·To re-calibrate

·To use right voltage

·To check error of four

corners

·To check load cell

Stay or show

wrong

symbols

·Short distracting outside

·To reboot or re-plug in

power

·To

reboot

after

30

minutes

Show “

·

” on

the left all

long

or

repeatedly

·Bad operation environment

(such as big air,vibration,big

change in room temperature

and so on.)

·To

control

working

temperature

Show

“No

Cou”

·No set constant before

counting

·constant is too big or small

·To set average value

constant

·To re-calibrate

No data

·Distracted (by static… ) and

crash

·To reboot or re-plug in

power

No back to

zero

·Wires loose

·To check all wires

No

calibration

·PCB is broken

·Load cell is broken

·Wires loose

·To change PCB

·To check load cell

·To check all wires

No Backlight ·Backlight is broken

·To change new backlight

Reviews: