12. Accessories:
Scale Main Frame························································· 1unit
Scale Pan·······································································1unit
User’s Manual······························································1unit
AC Adapter···································································1unit
OPT:
Calibration Weight ·····························································1unit
Rechargeable Battery··························································1unit
1. Environment Conditions
1.1. Place the balance in stably and smooth away vibration, sunlight, airflow
and strong electromagnetic wave .
1.2. Temperature
:
5
C
~
35
C
Temperature Fluctuation
:
≤5
C/h
Humidity
:
50
~
85
2. Main Technical Specifications
Model
DW1003L DW1503L DW1002L DW2002
L
DW3002L
Capacity
0-100 kg
0-150 kg
0-100 kg
0-200 kg
0-300 kg
Accuracy
1 g
1 g
10 g
10 g
10 g
Tare Range
0-100 kg
0-150 kg
0-100 kg
0-200 kg
0-300 kg
Repeatabilit
y
< = 3 g
< = 5 g
< 10 g
< 20 g
< 20 g
Linear
< = 5 g
< = 8 g
< 10 g
< 20 g
< 30 g
Stable time
3 s
Pan
500
×
400 mm
LCD
indicator
250 mm
×
150 mm
Outline
500
×
400
×
940 mm
Net Weight
11 kg
Power
110-240V/50-60HZ
Cal Weight
60 kg
100 kg
50 kg
100 kg
200 kg
3. Operation
3.1. Plug in the electronic balance and warm up 10 minutes is better.
3.2. Turn-on. It displays “8.8.8.8.8.8.”, “Maximum capacity”,”S-CAL” in turns.
Under weighing mode, it shows “0”, “0.0”,”0.00” or “0.000”.
4. Calibration
4.1. Single-range calibration
Press
for 3 seconds
Show “CAL*** “
“g” flicker
Put a
standard weight on the pan
Show “-CAL- “
Show weight value
Enter
into weighing mode. After calibration, if you found it still shows not exactly,
please re-calibrating.
4.2. Mult-range calibration
Press
for 3 seconds
Press “CAL” again and again, it will show
different weight value and “CAL L”, you can choose one kind of weight
values to calibrate on the basis of single-range calibration, or you also can
choose “CAL L” to perform the linear calibration.
When the screen shows “no CAL”, it means you didnot put weight on the pan
in 10 seconds. It will leave calibration mode automatically.
5. Weighing
5.1. Read the value on the screen after the black spot what is at the bottom
left disappeared. (That means the value is stable)