23
EN
The moisture measurement result is presented taking into
account the corrections connected with the temperature
of the sample (automatic temperature compensation),
g) when the measurement is finished, untwist the tighten-
ing cap and empty the measuring chamber You need
to do it carefully so that not to cause any mechanical
damage to the temperature sensor which is at the bot-
tom Since the measuring chamber is empty, the device
is ready for further operation In order to perform meas-
urements of another sample of the same species, fill the
measuring chamber again and twist the cap proper-
ly Then start the measurement with the
button
(or with the help of the
button go back to the list
to select another species)
NOTICE!
Proper cleaning of the measuring chamber
is very important when you change the species of
the sample and when you measure the samples with
different and increased humidity,
h) the final result of the moisture of the given sample
is the average of, let’s say, last three measurements
The TG pro calculates it automatically and shows on the
left side of the display (below the temperature result),
for example:
TG pro
21.6 0C
9.7
%
9.9%
Canola
Esc Clr
Start
The number of measurements from which the average is
calculated can be changed in the main menu The results
above and below the range are not taken into account
when the average is calculated If you want to calculate
the average result from the beginning, you can clear it
(function
Clr
) with the help of the
button or go back
to the list of the species with the help of the
button,
i) The Twist Grain pro enables saving of 50 thousands of
measurements to the internal memory of the device with
the help of the
button (function
Mem
) In order to use
it, go to the main menu and change to ”advanced” mode
Additionally, the user can choose if they want to save the
current result of the average of the last measurements
CHAPTER 6 | HUMIDITY MEASUREMENTS
Summary of Contents for Twist Grain pro
Page 3: ...2...
Page 4: ...3 EN INTRODUCTION...
Page 6: ...5 EN EQUIPMENT CHAPTER 1...
Page 8: ...7 EN CONSTRUCTION OF THE DEVICE CHAPTER 2...
Page 10: ...9 EN CHAPTER 2 EXTERNAL CONSTRUCTION APPROXIMATE DIMENSIONS 7 cm 7 5 cm 16 5 cm...
Page 11: ...10 CHAPTER 2 EXTERNAL CONSTRUCTION...
Page 12: ...11 EN KEYBOARD FUNCTIONS CHAPTER 3...
Page 14: ...13 EN HOW TO START THE DEVICE CHAPTER 4...
Page 18: ...17 EN NOTES FOR MEASUREMENTS CHAPTER 5...
Page 21: ...20 CHAPTER 5 NOTES FOR MEASUREMENTS...
Page 22: ...21 EN HUMIDITY MEASUREMENTS CHAPTER 6...
Page 26: ...25 EN EXTERNAL TEMPERATURE PROBE MEASUREMENTS CHAPTER 7...
Page 29: ...28 CHAPTER 7 EXTERNAL TEMPERATURE PROBE MEASUREMENTS...
Page 30: ...29 EN LIST OF AVAILABLE SPECIES AND HOW TO ADD NEW SPECIES CHAPTER 8...
Page 32: ...31 EN UPDATE VIA USB CHAPTER 9...
Page 34: ...33 EN MODIFICATION OF HUMIDITY RESULTS CHAPTER10...
Page 37: ...36 CHAPTER 10 MODIFICATION OF HUMIDITY RESULTS...
Page 38: ...37 EN MAIN MENU CHAPTER11...
Page 50: ...49 EN CHANGE OF BATTERIES CHAPTER12...
Page 53: ...52 CHAPTER 12 CHANGE OF BATTERIES...
Page 54: ...53 EN FINAL NOTES CHAPTER13...
Page 56: ...55 EN TECHNICAL DATA CHAPTER14...
Page 59: ...58...