SECTION 14 | TECHNICAL DATA
Dimensions
16.5 × 7.0 × 7.5 cm
Unit weight
520 g (with a set of batteries and nut)
Sample loading
manual
Moisture measurement method
impedance, at 4 frequencies
Sample volume
90 ml
Measurement control
Single chip microcomputer
Power supply
4 batteries type AA, 1.5 V, 1 battery type CR2032, 3V
Power consumption
from 30 to 72 mA depending on user’s settings
Estimated life of battery set
77 hours at 0% backlighting
54 hours at 30% backlighting
Battery low indication
graphic
Measurement resolution
moisture content – 0.1%, temperature – 0.1°C/°F
Accuracy of moisture content measurement
± 0.5% for standardized grain
± 1% in the range up to 10 % of moisture content
± 1.5% in the range above 10 % of moisture content and can increase
with the increase ontent with the increase of sample’s moisture content
Accuracy of temperature measurement
± 0.5°C in the range of 0°C to 85°C / ± 0.9°F in the range of 32°F to 185°F
Range of temperature measurement
from 0 °C to 125 °C
Result modification
using keyboard – modification option
Data recording
internal memory
Internal memory capacity
– 50,000 measurements with date and time
– 20 external temperature probes automatically recognized by ID number
– 20 language versions
– 400 species
Summary of Contents for TG pro coffee
Page 3: ...2...
Page 4: ...3 EN INTRODUCTION...
Page 6: ...5 EN EQUIPMENT SECTION 1...
Page 8: ...7 EN DESIGN OF THE DEVICE SECTION 2...
Page 10: ...9 EN SECTION 2 DESIGN OF THE DEVICE APPROXIMATE DIMENSIONS 7 cm 7 5 cm 16 5 cm...
Page 11: ...10 SECTION 2 DESIGN OF THE DEVICE...
Page 12: ...11 EN KEYBOARD FUNCTIONS SECTION 3...
Page 14: ...13 EN STARTING THE DEVICE SECTION 4...
Page 18: ...17 EN NOTES TO THE MEASUREMENTS SECTION 5...
Page 21: ...20 SECTION 5 NOTES TO MEASUREMENTS...
Page 22: ...21 EN HUMIDITY MEASUREMENTS SECTION 6...
Page 26: ...25 EN MEASUREMENTS WITH AN EXTERNAL TEMPERATURE PROBE SECTION 7...
Page 29: ...28 SECTION 7 MEASUREMENTS WITH AN EXTERNAL TEMPERATURE PROBE...
Page 30: ...29 EN LIST OF AVAILABLE SPECIES AND ADDING NEW SPECIES SECTION 8...
Page 32: ...31 EN UPDATE VIA USB SECTION 9...
Page 34: ...33 EN MODIFICATION OF HUMIDITY RESULTS SECTION10...
Page 37: ...36 SECTION 10 MODIFICATION OF HUMIDITY RESULTS...
Page 38: ...37 EN MAIN MENU SECTION11...
Page 50: ...49 EN BATTERY REPLACEMENT SECTION12...
Page 53: ...52 SECTION 12 BATTERY REPLACEMENT...
Page 54: ...53 EN FINAL REMARKS SECTION13...
Page 56: ...55 EN TECHNICAL DATA SECTION14...
Page 59: ...58 SECTION 14 TECHNICAL DATA...