Draminski TG pro coffee Manual Download Page 48

47

EN

 

Green Coffee Arabica

Value of change

-6.0

6.0

0.0%

7 . 

Add species 

First, please refer to the following section: LIST OF AVAILA-
BLE SPECIES AND ADDING NEW SPECIES .
To add a new species in the TG pro coffee moisture meter, 
go to the 

Main menu

 using the 

 key, then use the 

 

or 

 key to select the 

Add species 

option and confirm 

with the 

 key, e .g .

 

08.10.2018

Main menu

Temperature measureme

Settings

Result modification

Add species

  

08.10.2018

Enter code:

0

 1 2 3 4 5 6 7 8 9

A B C D E F

To enter a special code, select the appropriate characters 
with the 

 

or 

 key and confirm with the 

 key (to 

delete the character, select the “

<

” symbol and confirm 

with the 

 key) .

When you enter the whole code, the “

Accept code?

” mes-

sage appears which must be confirmed with the 

 key, 

e .g .

 

Accept code?

08.10.2018

Enter code:

NOTE: 

By entering the same code, the species are added 

and removed (alternately) in the available device list . The 
codes are matched to the device, so if, for example, you 
want to activate a new species in two TG pro coffee mois-
ture meters, you will need two different codes .

8 . 

About 

To check the device information and manufacturer’s con-
tact details, go to the 

Main menu

 using the 

 key, then 

use the 

 

or 

 key to select the 

About 

option and 

confirm with the 

 key .

SECTION 11 | MAIN MENU

Summary of Contents for TG pro coffee

Page 1: ...www draminski com ISO 9001 Drami skiTGprocoffee Smart moisture meter with sample compression for coffee beans and cocoa MANUAL EN...

Page 2: ...S TO THE MEASUREMENT 17 SECTION 6 HUMIDITY MEASUREMENTS 21 SECTION 7 MEASUREMENTS WITH AN EXTERNAL TEMPERATURE PROBE 25 SECTION 8 LIST OF AVAILABLE SPECIES AND ADDING NEW SPECIES 29 SECTION 9 UPDATE V...

Page 3: ...2...

Page 4: ...3 EN INTRODUCTION...

Page 5: ...time reserves the right to introduce changes and improvements in design and soft ware DRAMI SKI S A also reserves the right to amend the contents of the manual Read this manual carefully before starti...

Page 6: ...5 EN EQUIPMENT SECTION 1...

Page 7: ...2 Measuring chamber clamping nut with mounted 3V CR 2032 type battery 3 Transport case 4 Case strap for hanging 5 USB cable for communication with a computer 6 Instruction manual 7 4 x 1 5 V AA LR6 t...

Page 8: ...7 EN DESIGN OF THE DEVICE SECTION 2...

Page 9: ...ality ABS 4 Graphic LCD display with LED backlighting 5 Diaphragm keyboard with 4 keys 6 Mini USB socket with rubber cover 7 Compartment for four 1 5 V AA LR6 type batteries 1 Nut housing made of poly...

Page 10: ...9 EN SECTION 2 DESIGN OF THE DEVICE APPROXIMATE DIMENSIONS 7 cm 7 5 cm 16 5 cm...

Page 11: ...10 SECTION 2 DESIGN OF THE DEVICE...

Page 12: ...11 EN KEYBOARD FUNCTIONS SECTION 3...

Page 13: ...rning on the main menu by holding down the key for 2 seconds Cancelling program functions Approval of the program functions Start of the measurement Navigation through the menu Setting values in menu...

Page 14: ...13 EN STARTING THE DEVICE SECTION 4...

Page 15: ...advanced mode you can download from our website at www draminski com a special software for communication with your computer which allows you to download data from TG pro coffee to your computer creat...

Page 16: ...e standby mode for a few min utes the power will be off automatically this time can be changedinthemenu seesection MAINMENU Thedevice willsignalthisbyactivatingthebacklightandthedisplaywill show a cou...

Page 17: ...sc function returns to the list of available species Clr function clears the average result Mem function saves the result to the internal memory Start function starts the next measurement 08 10 2018 1...

Page 18: ...17 EN NOTES TO THE MEASUREMENTS SECTION 5...

Page 19: ...ighten the nut clean the thread regularly without using sharp tools When cleaning the chamber remember to do it carefully so as not to damage the temperature sensor located at the bottom of the measur...

Page 20: ...verestimating or underestimating the results by a similar value within the whole measurement range in relation to the laboratory method moisture balance a result modification should be applied see sec...

Page 21: ...20 SECTION 5 NOTES TO MEASUREMENTS...

Page 22: ...21 EN HUMIDITY MEASUREMENTS SECTION 6...

Page 23: ...t onto the measuring chamber s thread until you hear the buzzer emitting a beep sound coming from the nut This means the correct compres sionforce buttosavethebatteryofthebuzzer unscrew the nut as lit...

Page 24: ...e the average of e g the last 3 measurements TG pro coffee calculates it automatically and displays it on the left side of the display above the temperature result e g 21 60C 9 7 9 9 Green Coffee Arab...

Page 25: ...ee Arabica Esc Mem Start j after completing the measurements in order to save the power source turn off the device by holding down the key for 5 seconds or using the Turn off option available in the m...

Page 26: ...25 EN MEASUREMENTS WITH AN EXTERNAL TEMPERATURE PROBE SECTION 7...

Page 27: ...the device s memory changing the temperature scale 0 C 0 F therefore to learn all the functions we recommend read ing section 11 MAIN MENU MAIN MENU 1 Tousethetemperatureprobe turnontheadvancedop erat...

Page 28: ...1 00 21 70C PROBE NO 1 Menu Mem Special software for data transmission is available at www draminski com under Products Moisture meters Drami ski TG pro coffee SECTION 7 MEASUREMENTS WITH AN EXTERNAL...

Page 29: ...28 SECTION 7 MEASUREMENTS WITH AN EXTERNAL TEMPERATURE PROBE...

Page 30: ...29 EN LIST OF AVAILABLE SPECIES AND ADDING NEW SPECIES SECTION 8...

Page 31: ...an activate an additional species by entering a special code in the main menu through the Add species option see section MAIN MENU In order to get a special code activating the species you need in you...

Page 32: ...31 EN UPDATE VIA USB SECTION 9...

Page 33: ...on your device and many more Check the availability of new updates at www draminski com under Products Moisture me ters Drami ski TG pro coffee Our programmers spared no effort to make the update extr...

Page 34: ...33 EN MODIFICATION OF HUMIDITY RESULTS SECTION10...

Page 35: ...thod moisture balance and not to other moisture meters because they may be misleading The Drami ski TG pro coffee moisture meter allows you to modify correct the readings yourself When making modifica...

Page 36: ...35 EN A detailed description of how to modify the humidity results and how to restore the factory settings can be found in the section MAIN MENU SECTION 10 MODIFICATION OF HUMIDITY RESULTS...

Page 37: ...36 SECTION 10 MODIFICATION OF HUMIDITY RESULTS...

Page 38: ...37 EN MAIN MENU SECTION11...

Page 39: ...to this function the user can quickly and conven iently turn off the device without the necessity to hold down the key for 5 seconds and wait for the power auto off option 2 Species list To return to...

Page 40: ...i USB socket go to the Main menu then using the or key select the Temperature measurement option and confirm with the key NOTE To use the option of saving the results to the de vice s memory add the p...

Page 41: ...lect the appropriate value using the or key and confirm with the key e g 08 10 2018 Display Backlight intensity Backlight off time LCD contrast 08 10 2018 0 100 30 Backlight intensity 5 2 b Backlight...

Page 42: ...iate value using the or key and confirm with the key e g 08 10 2018 Settings Language Display Power auto off Number of averages 08 10 2018 Power auto off 1 30 3 min 5 4 Number of averages The device a...

Page 43: ...r auto off Number of averages Temperature scale 08 10 2018 Temperature scale Celsius Fahrenheit 5 6 Date time TG pro coffee has a real time clock so that the measure ment results are stored together w...

Page 44: ...8 10 2018 Date notation Imperial Metric 5 6 d Time notation to change the format of date dis play go to the Main menu Settings Date time Time notation then use the or key to select the appropriate opt...

Page 45: ...name select the name you want to change with the arrows and accept it with the key enter a new name for the probe use the ar rows to change and the key to confirm the high lighted characters when the...

Page 46: ...ults Averages 5 8 b Delete humidity results to delete all stored hu midity results go to the Main menu Settings Memory then use the or key to select the Delete humidity results option and confirm with...

Page 47: ...ase or decrease the humidity results and confirm the choice with the key e g 08 10 2018 Main menu Operating mode Temperaturemeasureme Settings Result modification Choose species Cocoa Green Coffee Rob...

Page 48: ...acters with the or key and confirm with the key to delete the character select the symbol and confirm with the key When you enter the whole code the Accept code mes sage appears which must be confirme...

Page 49: ...ss and contact details of Drami ski S A on page 2 08 10 2018 Main menu Settings Result modification About Add species www draminski com TG pro coffee Firmware v 1 4C K 0 10 SN TC000030 DRAMI SKI S A u...

Page 50: ...49 EN BATTERY REPLACEMENT SECTION12...

Page 51: ...eries To replace the batteries Remove the battery compartment cover at the back of the case and remove the used batteries using the ribbon Place the ribbon crosswise inside the battery compart ment so...

Page 52: ...d disappears or is visibly weaker replace the battery with a new one To replace the battery Unscrew the two screws in the nut cover Lift the cover and gently lift the battery out of the base by liftin...

Page 53: ...52 SECTION 12 BATTERY REPLACEMENT...

Page 54: ...53 EN FINAL REMARKS SECTION13...

Page 55: ...om electrolyte leakage We recommend the use of batteries from reputable manufacturers In case of problems with the device or difficulties in in terpreting the results we recommend before sending the d...

Page 56: ...55 EN TECHNICAL DATA SECTION14...

Page 57: ...c Measurement resolution moisturecontent 0 1 temperature 0 1 C F Accuracy of moisture content measurement 0 5 forstandardizedgrain 1 intherangeupto10 ofmoisturecontent 1 5 intherangeabove10 ofmoisture...

Page 58: ...iew batterystatuspreview temperaturemeasurementwithanexternalprobe changeofoperatingmode basic advanced modificationofhumidityresults addingnewspeciesthroughspecialcodes calculationoftheaverageresult...

Page 59: ...58 SECTION 14 TECHNICAL DATA...

Page 60: ...www draminski com DRAMI SKI S A Owocowa 17 10 860 Olsztyn Poland phone 48 89 527 11 30 e mail wilgo draminski com instr TGproC0119EN1 1...

Reviews: