15
PREPARING THE DEVICE FOR OPERATION
Plugging in the probe
Warning!
Probe connector should be protected from mechanical damage, dirtying and wetting. Probe wire
should be protected from overburden, strong repeated bending and pulling.
After precise sticking in the round connector to the nest in the casing, one should screw probes in holding
them by the metal part in order to ensure the full and correct contact of the probe connector with the
nest in the casing.
Ultrasound scanner DRAMIŃSKI Sono is equipped in one multipin socket for connecting probes.
Warning!
Probes should be exchanged when the device is switched off.
In order to change the probe, one should unscrew the connector holding it by the metal part, take it out
and then stick the new probe in and screw the connector in properly.
The pinned up head will automatically be recognized by the system and the sector of depicting will
change appropriately to the type of the connected probe.
Probe correctly pinned up to the device
Pinning up and checking the state of charging the battery pack
In order to check the state of charging the batteries, one should attach the pack to the device and
switch it on. A color indicator displayed on the information bar will inform the user about the time the
device could work using the energy stored in the battery pack.
I
--
I
red color indicator – the pack needs charging, only 15 minutes work possible
I
I
yellow color indicator – around 1.5 hour work possible
I
I
green color indicator – minimum 2 hour work possible, 6 hours with a fully charged
battery pack