Supplement Zeus
Infinity Empowered
15
English
Supplementary information
Breathing system mount
Prerequisite:
– The breathing system has been removed.
Perform the following steps:
1
Use a cloth dampened with disinfectant or a
disposable cloth that is ready for use.
2
Perform the surface disinfection of the
breathing system mount by wiping it back and
forth at least 3 times. In particular, disinfect the
optically highlighted connections (A).
Repeat steps 1 and 2 four times respectively. Use
a new cloth for each repetition of steps 1 and 2.
073
NOTE
– Do not use any alcohol-based disinfectants.
– Remove deposits.
– Ensure lint-free cleaning and disinfection.
– Do not allow fluid to get inside.
WARNING
Risk due to penetrating liquid
Penetrating liquid may cause the following:
– Damage to the device
– Electric shock
– Device malfunctions
Ensure that no liquid penetrates the device.
A (6x)