
14
Dräger X-dock 5300/6300/6600
Use
If an error occurs:
Remove from the module and re-insert.
If this does not correct the error, have the module repaired.
Overview of charge status LEDs
5.4
Station self-test
A self-test is carried out:
When the station is started up.
When the last successful self-test was more than 24 hours
ago and a test is being carried out.
The station is tested for leaks, pump function, and the software
version of the individual modules and the master.
5.5
Conducting a test
The following tests are preconfigured:
1. If necessary, open the test gas cylinders.
2. If necessary, switch on the X-dock.
3. Perform a visual inspection of the gas measurement
systems (see chapter 5.1 on page 13).
4. Insert the instruments into the modules (see chapter 5.2 on
page 13).
If Single mode is activated:
The preset test is started automatically by closing the
module cover.
The Status LED flashes blue.
The individual test phases are displayed.
If Favorites mode is activated:
Select required test from favourites bar.
The test will be started automatically.
The Status LED flashes blue.
The individual test phases are displayed.
If Test scheduler mode is activated:
If necessary, log user out on the station (see chapter 4.5 on
page 8).
The preset test is performed according to the configured
time schedule.
If Log-in mode is activated:
Log user in on the station (see chapter 4.5 on page 8).
Select the desired test from the favourites bar.
The test will start automatically.
The status LED flashes blue.
The individual test phases are displayed.
Test passed:
Confirmation is shown on the display.
The Status LED flashes green.
If required, select the desired instrument field for additional
information.
Remove the instrument from the module.
CAUTION
A short-circuit in the charging contacts in the modules,
e. g. due to metal objects falling into the device, will not
damage the station but should be avoided due to a
potential risk of overheating and error displays on the
module.
Colour
Status
Meaning
green
on continuously
Charge status 100 %
green
flashing
Battery charging.
red
flashing
Charging error
WARNING
When using methane, propane or butane in the range
>100 %LEL, an exhaust hose (max. 10 m long) must
be connected to the exhaust outlet to ensure the
extraction of excess explosive gas.
NOTICE
Single mode is activated by default.
Several tests can be started and executed in parallel in
the Single mode.
A failure of a LED, horn or vibration test results in a
negative evaluation of the overall test, and thus in the
locking of the respective gas measurement system.
A test of the sensor reserve is only performed with
sensors that support this function. The results are
displayed under test details and give information about
the state of the sensor.
Test 1:
QT
Fast bump test including alarm testing
Test 2:
EXT
Extended bump test incl. zero-point check and
alarm testing.
Test 3:
CAL
Calibration, alarm test, fresh air flushing and
certificate.
!
!
i
i
X-am 5000
X-dock 5300
00
01033286.eps
Summary of Contents for X-dock 5300
Page 1: ...Dr ger X dock 5300 Dr ger X dock 6300 6600 Technical Manual i...
Page 2: ......
Page 25: ...Glossary Dr ger X dock 5300 6300 6600 25...
Page 26: ......
Page 27: ......