Basics
Dräger X-dock 5300/6300/6600
11
There is also the option of setting an "increased sensitivity"
option for propane and pentane. This option artificially
increases the sensitivity to calibrate the sensors so that these
have approximately a nonane sensitivity level (in other words,
a sensitivity level as if they were calibrated for nonane). More
information on the subject of cross-sensitivity calibration can
be found in the relevant sensor data sheets.
To select the test gas for X-am 2000:
1. Select
X-am 2000
.
2. Select the required test gas from the list.
The following selection is available:
Methane - CH
4
(default setting)
Propane - C
3
H
8
Pentane - PENT
For propane and pentane, the "Increased sensitivity"
(vapour sensitivity) option can also be activated.
3. Confirm selection with
OK
.
To set the fresh air input:
1. Select
Fresh air
.
2. Select pump (fresh air inlet; default setting) or compressed
air inlet.
3. Confirm selection with
OK
.
The following settings can be made in
Expert settings
:
Ignore maximum concentration for quick bump test
Set test behaviour for missing test gases
To ignore the max. allowable concentration for the quick bump
test recommended by Dräger:
1. Select
Ignore max. conc. for BTQ
.
2. Activate checkbox (default setting: deactivated).
3. Confirm selection with
OK
.
If this function is activated, higher test gas concentrations can
be used for the fast bump test than those recommended by
Dräger.
To set the test behaviour with missing test gases:
1. Select
Missing gas
.
2. Activate checkbox (default setting: activated).
3. Confirm selection with
OK
.
This function can be used to set whether or not a test or
calibration is carried out when a required test gas is not
connected.
To select the test gas for PAC 7000 OV:
1. Select
Pac 7000 OV
.
2. Select the required test gas from the list.
The following selection is available:
Ethylene oxide - EO (default setting)
Carbon monoxide - CO
3. Confirm selection with
OK
.
To set the gas cylinder level indicator:
1. Select
Gas level monitoring
.
2. Activate or deactivate the
Gas level monitoring on
check
box.
3. Confirm selection with
OK
.
To reset the gas cylinder level indicator for a new test gas
cylinder:
1. Connect a new test gas cylinder to a test gas connection.
2. Select
>
Station gas configuration
.
3. Select desired gas inlet.
Select
Further options
and select to reset the gas cylinder
level indicator.
4.8
Managing tests
Select
>
Define test
.
The test overview will be displayed.
4.8.1
Selecting the test mode
The following test modes are available. The test modes define
the behaviour of the system when the user is logged off. The
Test scheduler and Log-in mode test modes are configurable.
NOTICE
The corresponding gas must be connected to one of
the gas inlets and set in the gas configuration.
WARNING
Only trained and experienced personnel are permitted
to activate this function, because an incorrectly
selected test gas concentration may result in a positive
test result even though the instrument alarmed too
late.
i
i
!
WARNING
If this function is deactivated, the corresponding
channel is not tested or calibrated.
NOTICE
The gas cylinder level indicator is only available for
cylinders that are configured via a Dräger part no.
Select test mode
►
◄
Configure test mode
Select test 1
►
◄
Edit test 1
Select test 2
►
◄
Edit test 2
Select test 3
►
◄
Edit test 3
!
i
i