Dräger X-dock 5300 Technical Manual Download Page 11

Basics

Dräger X-dock 5300/6300/6600

11

There is also the option of setting an "increased sensitivity"

option for propane and pentane. This option artificially

increases the sensitivity to calibrate the sensors so that these

have approximately a nonane sensitivity level (in other words,

a sensitivity level as if they were calibrated for nonane). More

information on the subject of cross-sensitivity calibration can

be found in the relevant sensor data sheets.

To select the test gas for X-am 2000:
1. Select 

X-am 2000

.

2. Select the required test gas from the list. 

The following selection is available: 

Methane - CH

4

 (default setting)

Propane - C

3

H

8

Pentane - PENT

For propane and pentane, the "Increased sensitivity"

(vapour sensitivity) option can also be activated.

3. Confirm selection with 

OK

.

To set the fresh air input:
1. Select 

Fresh air

.

2. Select pump (fresh air inlet; default setting) or compressed

air inlet.

3. Confirm selection with 

OK

.

The following settings can be made in 

Expert settings

:

Ignore maximum concentration for quick bump test

Set test behaviour for missing test gases

To ignore the max. allowable concentration for the quick bump

test recommended by Dräger:
1. Select 

Ignore max. conc. for BTQ

.

2. Activate checkbox (default setting: deactivated).
3. Confirm selection with 

OK

.

If this function is activated, higher test gas concentrations can

be used for the fast bump test than those recommended by

Dräger.

To set the test behaviour with missing test gases:
1. Select 

Missing gas

.

2. Activate checkbox (default setting: activated).
3. Confirm selection with 

OK

.

This function can be used to set whether or not a test or 

calibration is carried out when a required test gas is not 

connected. 

To select the test gas for PAC 7000 OV: 
1. Select 

Pac 7000 OV

.

2. Select the required test gas from the list. 

The following selection is available: 

Ethylene oxide - EO (default setting)

Carbon monoxide - CO

3. Confirm selection with 

OK

.

To set the gas cylinder level indicator:

1. Select 

Gas level monitoring

.

2. Activate or deactivate the 

Gas level monitoring on

 check

box.

3. Confirm selection with 

OK

.

To reset the gas cylinder level indicator for a new test gas

cylinder:
1. Connect a new test gas cylinder to a test gas connection.
2. Select  

Station gas configuration

.

3. Select desired gas inlet.

Select 

Further options

 and select   to reset the gas cylinder

level indicator.

4.8

Managing tests

Select  

Define test

.

The test overview will be displayed.

4.8.1

Selecting the test mode

The following test modes are available. The test modes define

the behaviour of the system when the user is logged off. The

Test scheduler and Log-in mode test modes are configurable.

NOTICE

The corresponding gas must be connected to one of

the gas inlets and set in the gas configuration.

WARNING

Only trained and experienced personnel are permitted

to activate this function, because an incorrectly

selected test gas concentration may result in a positive

test result even though the instrument alarmed too

late.

i

i

!

WARNING

If this function is deactivated, the corresponding

channel is not tested or calibrated.

 

NOTICE

The gas cylinder level indicator is only available for

cylinders that are configured via a Dräger part no.

Select test mode

Configure test mode

Select test 1

Edit test 1

Select test 2

Edit test 2

Select test 3

Edit test 3

!

i

i

Summary of Contents for X-dock 5300

Page 1: ...Dr ger X dock 5300 Dr ger X dock 6300 6600 Technical Manual i...

Page 2: ......

Page 3: ...aging instrument profiles 13 5 Use 13 5 1 Conducting a visual inspection 13 5 2 Inserting or removing the gas measuring device into or from a module 13 5 3 X am 125 module with charging function optio...

Page 4: ...re carried out by Dr ger Only trained and competent personnel are permitted to inspect repair and service the product as detailed in these Instructions for Use Use only genuine Dr ger spare parts and...

Page 5: ...ck 5300 6300 6600 5 2 Description 2 1 Product overview 0 X dock 5300 X am X dock 6300 6600 X dock 5300 Pac X dock 6300 6600 X dock 5300 6300 X dock 6600 max 10 00133286 eps 1 2 3 4 5 8 6 7 13 16 17 19...

Page 6: ...has been published under GPL LGPL or another open source licence They include GPL GPLv2 LGPL MIT PostgreSQL Apache Apache 2 zlib The source texts of the software can be obtained from Dr ger on a CD u...

Page 7: ...apter 4 7 on page 9 4 Change language if necessary a Select System configuration Language b Selected the required language c Confirm selection with OK 5 Set date and time where necessary a Select Sync...

Page 8: ...logon logoff Instrument field Favourites bar Selected test User logon logoff Instrument field Menu Select this button to access the menu see chapter 6 on page 16 Confirm Select this button to confirm...

Page 9: ...lly in addition 2 If necessary select Further options and to reset the gas cylinder level indicator 3 If required configure other test gas inlets in the same way Enter last name Enter first name Enter...

Page 10: ...are also used for calibration and testing These test gases are carbon monoxide CO and ethylene oxide EO For the X am 2000 it is possible to choose between three different test gases that are also use...

Page 11: ...th OK This function can be used to set whether or not a test or calibration is carried out when a required test gas is not connected To select the test gas for PAC 7000 OV 1 Select Pac 7000 OV 2 Selec...

Page 12: ...ters of the test name are displayed in the list of favourites Enter test name Select gas test Further options Further options i i No gas test only the activated options are run Quick bump test Test fo...

Page 13: ...the instrument 5 3 X am 125 module with charging function optional It is only possible to charge the instrument batteries using the X am 125 module with charging function The charging time is approx 4...

Page 14: ...ct the desired test from the favourites bar The test will start automatically The status LED flashes blue The individual test phases are displayed Test passed Confirmation is shown on the display The...

Page 15: ...ed If required select the desired instrument field for additional information Identify and rectify the error Repeat the test if necessary Overview of status LEDs 5 6 After use 1 If required remove ins...

Page 16: ...sion levels 5 i i Define test Station gas configuration System configuration Date time Data management Language Network System address Database DBMS address Info screen Update Service interval Devices...

Page 17: ...ean up 2 Select Start clean up 3 Confirm deletion of the database entries with OK The progress of the deletion is displayed 6 4 5 Displaying info screen 1 Select System configuration Info screen The f...

Page 18: ...required certificates from the list 4 Confirm selection with OK The selected certificates are copied onto the USB data storage device 6 5 6 Mobile synchronization This function can be used to copy the...

Page 19: ...Permission level 1 2 3 4 5 Perform preset tests in logged out status X X X X X Perform preset tests in logged in status X X X X Change password X Define test X Export data memory station X Set Station...

Page 20: ...nnections incorrect Test gas cylinder empty or closed Test gas pressure is too high Pressure regulator valve incorrectly set Check setting of pressure regulator valve Alarm element test not carried ou...

Page 21: ...on the USB storage device will be displayed 4 Select the desired firmware update from the list The selected firmware update is marked in blue 5 Start the firmware update with OK The progress of the i...

Page 22: ...played position markers as they appear 9 6 Cleaning The device does not need any special care If very dirty the equipment can be carefully wiped down with a damp cloth Carefully dry the device using a...

Page 23: ...e 2004 108 EC Name and description Order No Dr ger X dock 5300 X am 125 83 21 880 Dr ger X dock 5300 Pac 83 21 881 Dr ger X dock 6300 Master 83 21 900 Dr ger X dock 6600 Master 83 21 901 Dr ger X dock...

Page 24: ...ration DB Database DBMS Database Management System DL Data logger FAV Favourite HORN Horn LED Light emitting diode LEL Lower Explosive Limit MSD Mass Storage Device MST Master SPAN Span calibration SW...

Page 25: ...Glossary Dr ger X dock 5300 6300 6600 25...

Page 26: ......

Page 27: ......

Page 28: ...valstra e 1 23560 L beck Germany Tel 49 451 882 0 Fax 49 451 882 20 80 www draeger com 90 33 286 TH 4634 600 en Dr ger Safety AG Co KGaA Edition 07 November 2013 Edition 01 August 2012 Subject to alte...

Reviews: