Bare copper wire, class 2 according to IEC 60228 / VDE 0295
Values include a safety margin of 10%.
6.2 Installation
of
EC remote sensors
6.2.1 Installation
1. Insert the EC sensor in the remote sensor enclosure.
2. Attach the remote sensor enclosure to one of the following options.
3. Observe the preferred position of the sensor: downwards.
4. If the wall mount bracket is used, reinstall bayonet ring over the sensor, and turn
clockwise until sensor is locked in place.
5. Tighten adjusting screw.
Wall Mount Bracket
Fig. 6
Dimensions of Wall Mount Bracket
AWG
21
20
18
16
13
mm
2
0.5
0.75
1
1.5
2.5
122°F
50°C
feet
43
48
55
66
86
m
6
9
13
19
30
68°F
20°C
feet
45
50
57
70
93
m
7
10
14
21
34
Option
Mounting on a flat surface
Wall Mount Bracket 4544213
1)
1) Wall Mount Bracket is shipped with the EC remote enclosure.
Mounting to a bar or beam
Wall Mount Bracket 4544213
Mounting in a tube/duct
– Duct adapter 8317617
– Process adapter 8323404
41042
h
a
a
b
b
c
d
e
f
g
a 6.5 mm / 0.25‘‘
b 15 mm / 0.6‘‘
c 40 mm / 1.57‘‘
d 20 mm / 0.79‘‘
e 45 mm / 1.77‘‘
f 75 mm / 2.95‘‘
g 100 mm / 3.94‘‘
h M5 4 mm / 11/64‘‘