Supplement to the instructions for use
|
Isolette 8000 plus
13
Reprocessing
2.6 Validated
reprocessing
procedures
2.6.1
Overview of the reprocessing procedures of the components
2.6.2
Surface disinfection with cleaning
Components:
– Device surface including controller screen
– Trolley, rail, and holders
– Hood and walls
– Gaskets and grommets
– Removable parts of the main system
Surface disinfection using the following surface disinfectant is suitable for the
device and all components:
Components
Surface disin-
fection with
cleaning
Manual clean-
ing followed by
disinfection by
immersion
Machine
cleaning
with ther-
mal disin-
fection
Description of
the procedure
Device surface
including con-
troller screen
Yes
No
No
(see "Surface
disinfection
with cleaning",
page 13)
Trolley, rail,
and holders
Yes
No
No
(see "Surface
disinfection
with cleaning",
page 13)
Hood and walls Yes
No
No
(see "Surface
disinfection
with cleaning",
page 13)
Gaskets, grom-
mets
Yes
No
No
(see "Surface
disinfection
with cleaning",
page 13)
Removable
parts of the
main system
Yes
No
No
(see "Surface
disinfection
with cleaning",
page 13)
Humidity sys-
tem (if
installed)
No
No
Yes
(see "Machine
cleaning with
thermal disin-
fection",
page 16)
Surface disinfec-
tant
Manufacturer
Concentration
Contact time
Dizmozon plus
BODE Chemie
1.6 %
15 min