16
Supplement to the instructions for use
|
Isolette 8000 plus
Reprocessing
3. Be sure to clean all holes and indentations; then dry with a clean cloth or paper
towel.
Reprocessing the heater radiator and fan impeller
WARNING
Risk of death or serious injury
Failure to clean the heater radiator and fan impeller could result in sufficient lint
build-up to reduce airflow. Reduced airflow could affect the temperature control and
cause high oxygen concentrations.
►
Clean heater radiator and fan impeller.
CAUTION
Risk of injury or damage to the device
Damage can occur to the motor if liquid penetrates the area.
►
When cleaning the incubator shell, prevent liquids from entering the motor shaft
opening.
►
When cleaning the surface of the air circulation well, prevent liquids from
entering the motor shaft opening.
CAUTION
Risk of injury or damage to the device
Autoclaving can damage equipment or impact functional integrity.
►
Do not steam sterilize parts when disassembling for cleaning.
1. Remove any lint build-up on the heater radiator and fan impeller.
2. Wipe the heater assembly.
Do not immerse the heater assembly in liquid.
Reprocessing the air inlet filter chamber and cover
WARNING
Risk of death or serious injury
A dirty air inlet filter could affect performance or cause carbon dioxide (CO
2
)
buildup.
►
Ensure that the filter is checked on a routine basis commensurate with local
conditions and replaced as recommended in Maintenance on page 149.
Particularly, if the device is used in an unusually dusty environment, more
frequent replacements may be necessary.
Do not clean the air inlet filter.
1. Clean the filter chamber and its cover.
2. Install a new air inlet filter.
2.6.3 Machine
cleaning
with thermal disinfection
Use a washer-disinfector that meets the requirements of the standard ISO 15883.
Follow the manufacturer's instructions for the washer-disinfector.
Components:
– Humidity system