Adjusting working height
●
Connect to mains power supply.
1 Press right foot pedal to raise incubator.
2 Press left foot pedal to lower incubator.
●
Adjust to comfortable working height.
Hoses and cables must be the right length so that they
do not kink, tear or become squashed.
Do not store anything under the swivel cupboard.
Folding down back panel
●
Disengage back panel and fold down.
For instance, in order to
replace water bottles,
replace filter
connect climate sensor
●
Lift back panel up again and engage catch.
Connecting climate sensor
The climate sensor measures
air temperature
relative humidity and
O
2
concentration
●
Pull climate sensor out and take off.
●
Unscrew retaining ring, fit O
2
sensor capsule and
then screw retaining ring back into position.
[
Use two O
2
sensor capsules with the same expiry
date. Only use Dräger original sensor capsules.
See Order List on page 57.
●
Refit climate sensor and push back into place.
●
Disengage back panel and fold down.
●
Push sensor plug through slot between incubator
housing and back panel, from below.
3 Connect sensor plug to socket and screw safety
screws on firmly.
4 Route sensor cable through cable clips.
●
Lift back panel up again and engage.