Switching from manual mode to automatic mode:
1 Press
ç
key to switch humidity module off.
2 Green Control LED goes out.
1 Press
ç
key again to activate humidity module in
automatic mode.
2 Green Control LED is lit.
Alarm for water shortage:
3 Red Alarm LED flashes,
4 red H
2
O LED and
5 display flash.
Intermittent tone commences.
●
Replenish water supply, see p. 19.
The intermittent tone can be suppressed for 10 minutes:
6 Press
G
key,
7 yellow LED and
3 red Alarm LED are lit.
When cause of alarm has been rectified:
●
LEDs go out, intermittent tone ceases.
Other alarms, see "Fault, Cause, Remedy", p. 41 and 45.
Switching off humidity control:
8 Press
ç
key = humidity module switched off.
9 Green Control LED goes out.
Operation
Controlling humidity
30
% FEUCHTE/HUMID.
Control
%
%
1
2
Check
Control
%
%
3
6
4
5
7
% FEUCHTE/HUMID.
Control
%
%
8
9