Alarms
For deviations greater than
±
0.5
°
C between set value
and actual value of skin temperature:
1 Red Alarm LED and
2 Red
±
0.5
°
C LED flash,
3 display flashes and intermittent tone commences.
The intermittent tone can be suppressed for 10 minutes:
4 Press
G
key.
5 Yellow
g
LED and
1 red Alarm LED are lit.
When measured value is within
±
0.5
°
C of the set value
again,
2
±
0.5
°
C LED and
1 red Alarm LED go out.
intermittent tone ceases.
5 Yellow
g
LED goes out.
If the sensor plug is disconnected or sensor is faulty:
6 Display of 3 dashes in centre of display.
After 15 seconds:
●
Intermittent tone commences.
6 Display of 3 flashing dashes in centre of display.
7 Red Alarm LED and
8 red Sensor alarm LED are lit.
Then:
●
Connect sensor without delay or replace skin
temperature sensor.
The intermittent tone can be suppressed for 10 minutes:
●
Press
G
key, yellow
g
LED and red Alarm LED are
lit.