Dräger Incubator 8000 IC Instructions For Use Manual Download Page 19

Wait for warming-up time.
Wait for about 35 minutes before use.

Heat incubator in "air temperature control" operating
mode.

If incubator has to be used urgently, operate in standby
mode:

Set air temperature to 32 

°

C to 36 

°

C, see p.20.  

Oxygen control off, see p.33.

Humidity control off, see p.30.

Just before placing the baby in the incubator:

Fill water bottles

If required use protective gloves.

Open back panel and fold down. 

2  Disconnect empty water bottles and refill with 500 mL

clean, distilled or demineralised water (hospital
dispensary). Do not use any additives.

Refit filled water bottles and close back panel 
again.

Then disinfect hands.

Use 
– DIN 58363-IL 500-FL-BK1 or
– ISO 8536-1-IL 500-CL-HC1 Form A, 
infusion bottles only.

If a water container is used:

Fill water container

If required use protective gloves.

Take cover of water container off and

fill container to mark with clean, distilled or 

demineralised water (hospital dispensary).
Do not use any additives. Take care not to overfill.

Put cover of water container back.

Then disinfect hands.

Placing baby in the incubator

Open front door and pull out bed.

Put the baby on the bed and slide the bed back in
position.

Close the front door and check that the catches are
properly engaged.

Adjust the bed as required.

 
Adjusting working height

Adjust working height, as required, see p. 8.

19

Operation

1

2

1

2

Summary of Contents for Incubator 8000 IC

Page 1: ...D Incubator 8000 IC Instructions for Use Software 21 n 1 505 94...

Page 2: ...tions 17 Controlling air temperature 20 Using skin temperature control 23 ThermoMonitoring 28 Using humidity control 29 Using oxygen control 31 Supplying oxygen manually 34 Using electronic baby scale...

Page 3: ...apparatus is neither approved nor certified for use in areas where combustible or explosive gas mixtures are likely to occur Safe connection with other electrical equipment Electrical connections to e...

Page 4: ...controlled on the basis of the arterially measured oxygen partial pressure in the patient s blood If this is not done there is a risk of hyperoxaemia damage to the eyes and hypoxaemia damage to the br...

Page 5: ...ack until it rests on sloping surface of canopy It is then possible to attend to the patient If necessary Pull lower section down pulling gently against pivots to do so and lower until it rests on hou...

Page 6: ...lift up canopy Open front door Tip canopy back as far as it goes Taking out the bed Open front door Pull bed out forwards as far as it goes After completing care procedure push bed back as far as it w...

Page 7: ...heir greater heat production may cause the incubator temperature to rise and if so the double walls should be removed Open front door Disengage double walls from upper bracket and remove To put back F...

Page 8: ...te sensor The climate sensor measures air temperature relative humidity and O2 concentration Pull climate sensor out and take off Unscrew retaining ring fit O2 sensor capsule and then screw retaining...

Page 9: ...tighten Replace bed in the incubator and close front door Bronchial suction equipment See also relevant Instructions for Use To mount on incubator Screw holder to left or right end face of trolley at...

Page 10: ...move foot strips from Babylog Tilt Babylog forward by about 45 1 Insert front latches into slots in mounting plate 2 Lower Babylog insert the rear latches into slots in mounting plate and secure at ba...

Page 11: ...table to pillar and tighten screw Make sure that swivelling area is clear Swivel cupboard To move swivel cupboard on right of incubator to left or to fit an additional swivel cupboard Push Allen scre...

Page 12: ...llow Instructions for Use of equipment being used Monitor O2 concentration O2 flowmeter Attach O2 flowmeter to rail Fit hose to connector on O2 flowmeter and 2 to O2 connector underneath incubator Pus...

Page 13: ...able through one of the flexible hose seals Push sensor plug into socket on Oxydig until it clicks into place Oxygen distributor Attach oxygen distributor order no 2M 18 810 to rail Routing cables and...

Page 14: ...has no cracks or sharp chipped edges Check that the hinges and catches on the canopy are in proper working order Check that the cables and hoses are routed correctly and safely Check that two O2 sens...

Page 15: ...air filter Fold down back panel 3 Open filter cover plate downwards If a filter is fitted Remove filter and check fitting date label is on edge of filter If filter is more than 2 months old Replace wi...

Page 16: ...Checking LEDs displays and alarm sound 5 Press k key For about 2 seconds all LEDs are lit except the mains power failure LED the digital displays show 88 8 and alarm sound commences Thereafter display...

Page 17: ...gh heating power Temperature drops slowly because of good thermal insulation Additional external heat sources such as sunlight heat lamps spotlamps electric cushions should be avoided These increase t...

Page 18: ...mia damage to the eyes and hypoxaemia damage to the brain Temperature of breathing gas During ventilation the breathing gas hoses may be additionally heated by the heated air circulating in the incuba...

Page 19: ...ter bottles and close back panel again Then disinfect hands Use DIN 58363 IL 500 FL BK1 or ISO 8536 1 IL 500 CL HC1 Form A infusion bottles only If a water container is used Fill water container If re...

Page 20: ...g display of air temperature set value is 33 0 C 5 Press P or p key briefly to select this set value The display will remain continuously lit Setting values in 28 C to 37 C range The set value can be...

Page 21: ...7 C control LED will go out The lock against setting a higher range of values will be in force again The alarm limit for high temperature will return to 38 C again If a value below 37 C is set after t...

Page 22: ...mittent tone ceases When the incubator is switched on the alarm sound is automatically suppressed for 30 minutes during the warming up phase 4 red 1 5 C LED and yellow g LED and red Alarm LED are lit...

Page 23: ...tes between the measured value for air temperature and word SEt 4 Confirm with P or p key and set a new value Using skin temperature control When the skin temperature control option is installed Do no...

Page 24: ...kidneys The sensor must never be attached under the baby since it would then be measuring and controlling core temperature and not skin temperature Check regularly that the skin temperature sensor is...

Page 25: ...alue for skin temperature 5 The measured value for air temperature continues to be displayed The incubator offers an appropriate value depending on the situation Situation Set value Set value for air...

Page 26: ...he set value and actual value of skin tempera ture also has an influence on the control of air tempera ture inside the incubator The patient s skin temperature can change suddenly as a result of feedi...

Page 27: ...sor plug is disconnected or sensor is faulty 6 Display of 3 dashes in centre of display After 15 seconds Intermittent tone commences 6 Display of 3 flashing dashes in centre of display 7 Red Alarm LED...

Page 28: ...om adhesive pad and place skin temperature sensor on pad Attach sensor to the patient s extremeties preferably on the foot using adhesive pad Fix sensor cable in place with adhesive plaster Connect Wi...

Page 29: ...incubator alternate in display 5 Press P or p key briefly 3 the word Aut remains continuously lit 4 Measured relative humidity is displayed Activating humidity control in manual mode 6 Press P or p ke...

Page 30: ...ttent tone commences Replenish water supply see p 19 The intermittent tone can be suppressed for 10 minutes 6 Press G key 7 yellow LED and 3 red Alarm LED are lit When cause of alarm has been rectifie...

Page 31: ...se to terminal unit of medical gas pipeline system 1 Press key 2 Green Control LED is lit 3 Yellow Cal LED and 4 word CAL flash 5 Set value 21 flashes in display Calibrate O2 sensor within one minute...

Page 32: ...played Value must be set within one minute or the extended set value range will be cancelled If value is re set below 40 vol O2 the extended range of set values is automatically cancelled Alarms When...

Page 33: ...41 and 44 Switching off O2 control module when there is a fault in O2 control module Err display when there is a sensor alarm which cannot be rectified when the deviation from set value remains above...

Page 34: ...uired is reached switch off O2 supply at flowmeter Recommended set values O2 concentration Vol approx 25 30 35 40 45 50 55 60 O2 flow L min 1 5 3 5 5 5 7 5 9 11 13 15 Using oxygen limiter Preparation...

Page 35: ...may remain in incubator Bed can be tilted as before Using vacuum mattress Accessory The mattress can be formed into any shape and will keep this shape after the mattress has been evacuated In this way...

Page 36: ...w safety advice see p 18 Using BabyLink Incubator interface optional The Babylink Incubator interface provides incubator data actual values set values alarms via a serial interface At present measured...

Page 37: ...er bottles Fold down back panel Allow water to flow back into bottles 1 Remove bottles from holder and drain Hold connecting hose by cuff and detach from connector on heater Unscrew bottle holder Wate...

Page 38: ...iene and Microbiology list of surface disinfectants are used For medical and hygienic reasons disinfectants based on quaternary compounds should be used Damage may occur to the material of the incubat...

Page 39: ...te sensor Fan Seals Hose seals Connecting hose Bottle holder or water container Wash with detergent and rinse with clean water Disinfect in bath Allow time for the disinfection process then rinse with...

Page 40: ...hand port Testing Readiness for Operation see p 14 Allow water heater to run dry to disinfect heater to test that water shortage display is working properly Switch on humidity control Allow heater to...

Page 41: ...bator not working Call Dr gerService Red Inop LED is lit Fault in specific module Continous tone commences also Err display in a module except Skin temperature control module Switch off specific modul...

Page 42: ...up to 37 C lamps radiators sunlight Wait intermittent tone commences Air temperature above 40 C for incubator to cool down Range of values set up to 39 C Press key when temperature has dropped below...

Page 43: ...air temperature control Red Inop LED is lit in main module module while in operation Call Dr gerService Continuous tone commences Err is lit in actual value display Fault in the skin temperature Othe...

Page 44: ...e commences Yellow Cal LED flashes Request to calibrate O2 sensors at Calibrate O2 sensors CAL flashes in actual value display start and every 24 hours of Continuous tone commences continuous operatio...

Page 45: ...Check connection Err flashes in actual value display Fault in humidifier Switch off humidifier Red Inop LED is lit in main module Humidity control activated All other functions remain Continuous tone...

Page 46: ...ttery for replace once a year by power failure trained service personnel Disposal see below Inspection and every six months by trained service personnel Disposal of O2 sensor capsules and NiCd batteri...

Page 47: ...sted column and four castors 2 lockable 14 Pedals to adjust height Back view 15 Connector for climate sensor 16 Air filter 17 Back panel 18 Optional RS 232 connection 19 Mains cable 20 Bottle holder f...

Page 48: ...unction 1 5 Red N N N NLED is lit during mains power failure 1 6 Yellow g LED is lit when intermittent tone has been suppressed 1 7 Key to suppress intermittent tone for 10 minutes 1 8 Key to test fun...

Page 49: ...r temperature 2 7 Display for set value of air temperature left key to decrease set value right key to increase set value 2 8 Red 1 5 C LED flashes is lit when the actual value of the air temperature...

Page 50: ...Display for set value of skin temperature left key to decrease set value right key to increase set value 3 7 Red 0 5 C LED flashes is lit when the actual value of skin temperature deviates from the se...

Page 51: ...l value of oxygen concentration deviates from the set value by more than 5 vol 4 9 Red Sensor LED flashes when sensor is faulty or pulled out 4 10 Key to calibrate O2 sensor Humidity control module 5...

Page 52: ...erature drop at room temperature of 25 C and temperature inside incubator of 36 C Two hand ports opened 1 C Front door opened 5 C Increase of O2 oxygen concentration from 21 to 60 vol 7 min Humidifica...

Page 53: ...calibration Sensor subject to calibration IB 9 13 205 91 Calibration tolerances Display component 0 2 C subject to calibration Sensor subject to calibration 0 1 C Total measuring system 0 3 C Classif...

Page 54: ...rolled by a microprocessor When it is switched on and every 10 minutes thereafter the incubator automatically carries out a self test This test checks all modules which are particularly important for...

Page 55: ...ignal is given even when the flap with brief instructions is folded up If another alarm occurs when the alarm tone has already been suppressed the tone is automatically re activated Depending on what...

Page 56: ...d to connect peripheral equipment 83 06 488 BabyScreen 2M 21 707 ThermoView 2M 21 888 Name and Description Order No Photo Therapy 4000 2M 21 000 Trolley for phototherapy unit 2M 21 190 Bronchial sucti...

Page 57: ...r 2M 19 561 Screws to secure the bottle holder 2M 06 004 Skin temperature sensor not calibrated pack of 10 2M 21 726 Skin temperature sensor calibrated pack of 10 2M 22 010 Skin temperature sensor cal...

Page 58: ...ing sterilizing 38 Climate sensor connecting 8 Controlling air temperature 20 humidity 29 skin temperature 25 26 temperature inside incubator 17 Cooling incubator 17 Design 54 Doors ports and bed 5 Fa...

Page 59: ...function 26 connecting 24 peripheral connecting 28 placing 23 Standby 19 Stripping down 37 Suppressing intermittent tone 22 Swivel cupboard 11 Technical data 52 ThermoMonitoring 28 Tilting bed 6 15 V...

Page 60: ...C FR 00 054 with Serial No If no Serial No has been filled in by Dr ger these Instructions for Use are provided for general information only and are not intended for use with any specific machine or d...

Reviews: