Dräger DrugTest 5000 Technical Manual Download Page 25

Maintenance

Dräger DrugTest 5000 Test System

25

5

Maintenance

The Dräger DrugTest 5000 Analyzer is a sturdy device

requiring a minimum of maintenance and care. 

5.1

Maintenance intervals

We recommend that the unit be maintained on a regular basis

every 12 months.

Servicing only by Dräger Services or a Dräger authorized

service organisation.

We recommend signing a service contract with Dräger.

5.2

Quality Assurance Mode (QA mode)

The Dräger DrugTest Analyzer has the capability of analyzing

both positive and negative quality assurance (QA) cassettes. If

the QA cassette analyses do not provide positive results for

positive control cassettes and negative results for negative

control cassettes, the Analyzer will not perform any subject

testing until it has been serviced. If QA cassette analyses have

not been carried out on the Analyzer for more than 15 days, it

will not perform any further subject testing until such QA

cassette analyses have been completed.

To conduct quality assurance tests, see chapter 3.9

5.3

Cleaning

If necessary, wipe the outer surfaces of the unit with a damp,

soft cloth that has been dipped in a mild soap solution or a

commercial laboratory detergent (e.g. isopropanol).

The inside of the unit must not be cleaned by users.

5.4

Maintenance work

5.4.1

Troubleshooting and repair

If problems should arise during the use of the unit, which

cannot be solved by referring to the table “Troubleshooting” on

page 21, you should notify Dräger Services or a Dräger

authorised service organisation.

Some faults may require that the unit is returned for repair to

Dräger. For this case we have a repair substitution concept:

The Dräger Services in many cases has a repair-exchange

unit available to reduce down times.

1. The faulty unit is packed into the packing of the

replacement unit and the enclosed address label affixed.

2. Send the faulty unit to Dräger as soon as possible.

5.4.2

Battery maintenance

For optimal care of the battery, the unit should always be

connected to the mains via the charger. The unit automatically

ensures the best care of the battery (trickle charging).

Recharge the battery immediately after use and avoid storing

the unit with incompletely charged battery.

The Analyzer is equipped with an integrated lead storage

battery. It consumes a small standby current even if switched

off, which can discharge a fully charged battery in approx. 

2

months

. The battery capacity can be retained at an optimum

and for the long term if periods of low charge are avoided. 

To do so the following measures are required:

Always operate the Analyzer with the power pack if it is not

being used in mobile applications. A permanent connection

to the mains does not damage the integrated battery, but in

fact preserves its capacity.

For applications in vehicles use the 12V cable (order code

812166).

In mobile mode, if the unit indicates a low battery charge,

connect the battery charger and fully charge the battery as

soon as possible.

During storage always leave the battery charger

connected.

For long term storage of the unit where it is not possible to

connect the battery charger, fully charge the Analyzer

before placing it into storage and fully charge the battery

every 

six weeks

.

6

Disposal of electrical and 

electronic equipment

 EU-wide regulations for the disposal of electric

and electronic appliances which have been

defined in the EU Directive 2002/96/EC and in

national laws are effective since August 2005

and apply to this device. Common household

appliances can be disposed of using special

collecting and recycling facilities. However, as

this device has not been registered for

household usage, it must not be disposed of

through these means. The device can be returned to your

national Dräger Sales Organisation for disposal. Please do not

hesitate to contact the above if you have any further questions

on this issue.

CAUTION

Do not immerse the unit in liquids and make sure no

liquids enter into the connectors!

!

CAUTION

The built-in rechargeable battery will be damaged if the

unit is stored with a discharged battery.

!

Summary of Contents for DrugTest 5000

Page 1: ...r DrugTest 5000 Test System CA non IVD WARNING To properly use this product read and comply with these instructions for use i enus Technical Manual 3 fr Manuel technique 31 NOTICE For Forensic Use onl...

Page 2: ......

Page 3: ...ns for Use 10 3 3 Align the printer 10 3 4 Switch on Analyzer 10 3 5 Sampling 10 3 6 Analysing the sample 11 3 7 Possible results of the analysis 12 3 8 Switching Analyzer off 12 3 9 Menu 13 3 10 Conf...

Page 4: ...the unit makes any claims under guarantee void Annual inspections of the unit are recommended See Maintenance intervals on page 25 1 2 1 Not for use in areas of explosion hazard The unit is not approv...

Page 5: ...wer compartment cassette compartment 4 IRDA for printer communication and air outlet lateral 5 Carrying handle 6 Display 7 Air inlet 2 1 3 Rear side no illustration connections are labelled Socket for...

Page 6: ...state of charge of the integrated rechargeable battery Illustrations of the various charge state symbols see Explanation of Symbols on page 8 2 4 1 Mains operation Standard mode The Analyzer is conne...

Page 7: ...ive measuring system to detect searched substances or their metabolites which are contained in the sample in excess of certain concentration limits cut off and therefore delivers only a preliminary an...

Page 8: ...tative to perform a firmware update Display Charge status Keyboard lights are flashing Analyzer is switched off Tones Analyzer is switched off Mains connection established no signal one short tone con...

Page 9: ...additional accessories like for example printer keyboard bar code reader disposable gloves transport bag training test kits and hard case See Order list on page 27 Symbols Meaning Strictly follow the...

Page 10: ...h The Dr ger Mobile Printer automatically turns itself off after 4 minutes to save energy Therefore only turn the Dr ger Mobile Printer on just before printing or operate with the associated power pac...

Page 11: ...erforming the test A status bar on the display indicated the progress of the evaluation process Optional Depending on configuration enter the following data follow instructions on the display Enter te...

Page 12: ...n has been executed correctly will the Analyzer take this specific detection reaction into account In case of errors in the reaction process the affected detection reaction will not be evaluated to ex...

Page 13: ...n Use Confirm to continue logging in Use Skip to continue without logging in There are three options to log into the device as Standard Supervisor and Operator e g Service The menu is navigated by usi...

Page 14: ...steps Selection activated Functions which can be selected and activated are enabled by pressing the key OK Settings Language Language selection QA Mode Perform a quality assurance test Warning signal...

Page 15: ...age only after you end this function Quality assurance mode QA mode This feature allows you to conduct quality assurance QA tests Please follow the instructions on screen For more information see Chap...

Page 16: ...digits in the same way and acknowledge Once the last digit is acknowledged the Time and date submenu is displayed again Set date 1 Select the Date function 2 Use the keys or to set the first digit of...

Page 17: ...keyboard and press key 4 Save the entered user name by pressing the key 5 Following this enter a password for the new user press key and enter the password again for confirmation 6 Save the entered p...

Page 18: ...his function is used to display unit parameter versions 1 Use the key to select Versions and press the key to activate 2 End the function by pressing the key Device statistics This function is used to...

Page 19: ...se the key to select Measurement statistics and press the key to activate 2 End the function by pressing the key Measurement data This function is used to display a list of measurements in data memory...

Page 20: ...ces Further options available within the Diagnostics software may be subject of training and might be restricted to be performed by the Dr ger Services 1 Launch Dr ger Diagnostics Software 2 Select th...

Page 21: ...tem test of the device Measurement not possible Please carry out maintenance 10517 System test not successful Measurement is not possible Detailed inspection is necessary to find error Service is requ...

Page 22: ...he cassette has expired Measurement has been cancelled Remove cassette and cartridge and close the door 20006 Database failure Test cassette cannot be processed Update software Panel Type cannot be me...

Page 23: ...ssword Wrong password entered Please try again 22052 Change operator password new password already used Choose different password from existing one Invalid password Please choose another password 2205...

Page 24: ...re only sections of the indicator are blue or where the blue colouration is unevenly distributed If the collection of the oral fluid sample was carried out properly the test will typically take 30 sec...

Page 25: ...mal care of the battery the unit should always be connected to the mains via the charger The unit automatically ensures the best care of the battery trickle charging Recharge the battery immediately a...

Page 26: ...x D 200 mm x 250 mm x 220 mm Weight 4 5 kg including battery Temperature range Operation 5 C to 40 C 41 F to 104 F Storage Transport 20 C to 60 C 4 F to 140 F Interfaces optical IRDA printer PS 2 for...

Page 27: ...r 5 rolls 83 19 002 Carrier bag Dr ger DrugTest 5000 Analyzer 83 22 675 Transport case 83 25 333 Battery charger 12 V with mains cable for EU UK USA AUS 83 15 675 Vehicle power cable 12 V Power cable...

Page 28: ...placing both the test cassette and the cartridge into the Analyzer The Analyzer transfers the sample to the test cassette and the test development is initiated automatically Test The sample interacts...

Page 29: ...ood or other substances etc see also interfering substances Technical errors and procedural errors can also contribute to false results The taking of samples should be closely monitored If there is a...

Page 30: ...e help of GCMS or LCMS 9 2 2 Chromatography The term chromatography comprises physical methods whereby substances are separated by distribution between a static stationary and a moving mobile phase 9...

Page 31: ...ration l utilisation 38 3 3 Aligner l imprimante 38 3 4 Allumer l analyseur 38 3 5 chantillonnage 38 3 6 Analyse de l chantillon 39 3 7 R sultats possibles de l analyse 40 3 8 teindre l analyseur 40 3...

Page 32: ...ommandons une inspection annuelle de l appareil Voir Intervalles d entretien la page 53 1 2 1 Ne pas employer l o il y a un risque d explosion L utilisation de l appareil dans des endroits susceptible...

Page 33: ...artiment de la cassette 4 metteur IR pour communication avec l imprimante et vent lat rale 5 Poign e de transport 6 Affichage 7 Entr e d air 2 1 3 Arri re aucune illustration les branchements sont tiq...

Page 34: ...ir Explication des symboles la page 36 2 4 1 Utilisation avec branchement en mode standard Mode standard L analyseur est branch une source d alimentation externe Mode restreint L accumulateur interne...

Page 35: ...rmet de d tecter des substances donn es ou leurs m tabolites qui sont contenus dans l chantillon au del de certains seuils de concentration seuils et il ne donne donc qu un r sultat d analyse pr limin...

Page 36: ...ise jour du mat riel Affichage de l tat de charge Les t moins lumineux du clavier clignotent L analyseur est teint Tonalit s L analyseur est teint Branchement lectrique effectu aucun signal une br ve...

Page 37: ...ante un clavier un lecteur de code barres des gants jetables un sac de transport des trousses de d pistage pour formation et un bo tier Voir Liste de commandes la page 54 Symboles Signification Suivre...

Page 38: ...nutes d inactivit pour conomiser l nergie Par cons quent il ne faut allumer l imprimante qu au moment de s en servir avec le bloc d alimentation associ m me lorsqu on se sert de l accumulateur usage u...

Page 39: ...e test L cran affichera la progression du processus d valuation Facultatif Selon la configuration saisir les donn es ou suivre les directives affich es l cran Saisir les donn es du sujet Saisir les do...

Page 40: ...ut e correctement que l analyseur en tiendra compte En cas d erreurs du processus de r action la r action affect e ne sera pas valu e afin d exclure les r sultats erron s et elle sera qualifi e d inva...

Page 41: ...n d analyse Appuyer sur Confirmer pour poursuivre l identification Appuyer sur Sauter pour poursuivre sans s identifier Il est possible d utiliser l appareil en mode Standard Superviseur et Op rateur...

Page 42: ...uvent tre choisies et activ es sont identifi es en appuyant sur la OK R glages Langue Choix de la langue Mode d AQ Effectuer un test de contr le de qualit Signal d avertissement Activer et d sactiver...

Page 43: ...ode d AQ Cette fonction vous permet d effectuer des tests de contr le de qualit Veuillez suivre les instructions l cran Pour plus d informations voir le chapitre 5 2 Signal d avertissement Cette carac...

Page 44: ...rmer Apr s avoir confirm le dernier chiffre le sous menu Time and date Heure et date est affich de nouveau Fixer la date 1 Choisir la fonction Date 2 l aide des touches ou d terminer le premier chiffr...

Page 45: ...touche 5 Par la suite saisir un mot de passe pour le nouvel utilisateur appuyer sur la touche saisir le mot de passe de nouveau pour le confirmer 6 Sauvegarder le mot de passe du nouvel utilisateur e...

Page 46: ...ersions de param tres de l appareil 1 l aide de la touche choisir Versions et appuyer sur la touche pour l activer 2 Mettre fin la fonction en appuyant sur la touche Statistiques de l appareil Cette f...

Page 47: ...tre choix en appuyant sur la touche 2 Interrompre la fonction en appuyant sur la touche Donn es de mesurage Cette fonction permet d afficher une liste de tests dans la m moire des donn es avec la date...

Page 48: ...s d entretien Dr ger D autres options offertes par le logiciel Diagnostics peuvent faire l objet d une formation et tre r serv es aux services d entretien de Dr ger 1 Lancer le logiciel Dr ger Diagnos...

Page 49: ...n Communiquer avec votre repr sentant Dr ger Lors du test syst me de l appareil des d fauts se sont produits Mesurage impossible Veuillez effectuer 10517 chec du test syst me D pistage impossible Une...

Page 50: ...rmer le clapet 20005 Erreur de l usager Dur e de vie de la trousse expir e Utiliser une trousse valide La date limite de durabilit de la cassette a t d pass e Le mesurage a t annul Enlever la cassette...

Page 51: ...sse ant rieur Saisir un mot de passe correct Mot de passe incorrect Essayer nouveau SVP 22052 Changer le mot de passe d un op rateur le nouveau mot de passe est d j en utilisation Choisir un mot de pa...

Page 52: ...ique aux cas o seules certaines parties de l indicateur sont bleut es ou lorsque la coloration bleue n est pas distribu e galement Si l chantillonnage de fluides buccaux a t effectu correctement le te...

Page 53: ...e possible 5 4 2 Entretien de l accumulateur Pour un entretien optimal de l accumulateur l appareil devrait toujours tre branch son chargeur qui lui est branch l alimentation lectrique L appareil s as...

Page 54: ...approx 28 2 x 13 2 x 2 4 QWERTY clavier de configuration anglaise 83 15 497 AZERTY clavier de configuration fran aise 83 15 142 Imprimante mobile Dr ger 83 19 310 Papier pour imprimante mobile Dr ger...

Page 55: ...vertes d anticorps chantillons Les diff rentes trousses de d pistage ont t valu es pour en assurer la compatibilit avec les chantillons marqu s Il n est pas n cessaire de donner un traitement quelconq...

Page 56: ...age des substances transr actives ou des interactions non identifi es peuvent interf rer avec le test et provoquer de faux r sultats Les erreurs techniques et proc durales peuvent galement contribuer...

Page 57: ...ou liquide 9 2 2 Chromatographie Le terme chromatographie fait r f rence des m thodes physiques alors que les substances sont s par es par distribution entre une phase statique stationnaire et une pha...

Page 58: ......

Page 59: ......

Page 60: ...82 0 Fax 49 451 882 20 80 www draeger com 9100163 TH 4755 300 enUS_fr Dr ger Safety AG Co KGaA Edition 03 May 2021 Edition 01 June 2018 Subject to alteration Dr ger Safety AG Co KGaA Revalstra e 1 235...

Reviews: