background image

12

Instructions for use

  |  

Dräger BG ProAir

  

Use 

4.1.4 Reuse

Under the following prerequisites, the closed-circuit breathing 

apparatus can be used multiple times by the same user within 

24 hours:
– The officer-in-charge decides whether the remaining 

oxygen supply is adequate before each further use.

– The cylinder valve is fully closed after every use. The 

connecting piece is sealed with the protective cap.

– The compressed oxygen cylinder and CO

2

 absorber are 

not replaced.

– The ambient temperature is no less than +10 °C when the 

closed-circuit breathing apparatus is not being used.

– The closed-circuit breathing apparatus is placed in a 

upright position (not laid down).

4.2  Prior to first use

1. Perform a visual inspection. The closed-circuit breathing 

apparatus must be intact and complete.

2. Insert batteries into the closed-circuit breathing apparatus 

(see "Replacing the batteries", page 26).

3. Remove the thin, flexible protective foil from the LC 

display and the LED panel on the Connect ECU. The LC 

display and LED panel also have a clear, protective cover 

that must not be removed.

4. Carry out a leak and functional test (see "Testing the 

closed-circuit breathing apparatus", pag22).

4.3 Preparations 

for 

use

4.3.1 Freezing the ice packs (for apparatuses 

with ice cooler) 

1. Stack the ice packs flat on top of each other.
2. Freeze the ice packs at -16 °C for 24 hours.

If significant deformations are apparent on the ice pack, 

allow the ice pack to thaw and then freeze again. Dräger 

recommends using the respective freezing aid to prevent 

the ice packs from deforming.

Make sure that no ice is deposited on the ice packs. If 

necessary, place ice packs in a bag, seal the bag and 

freeze the ice packs whilst they remain in the bag.

Dräger recommends storing completely frozen ice packs 

in a sealed bag in the freezer.

4.3.2  Inserting ice packs (for apparatuses with 

ice cooler)

1. Open the closed-circuit breathing apparatus (see 

"Opening the closed-circuit breathing apparatus", 

page 16).

2. Open the ice cooler.
3. Insert the ice packs.

Only use ice packs that are not deformed. Do not press 

the ice packs into the ice cooler to ensure they do not 

become damaged and leak.

4. Close the ice cooler.
5. Close the closed-circuit breathing apparatus (see "Closing 

the closed-circuit breathing apparatus", page 22).

4.3.3  Donning the closed-circuit breathing 

apparatus

1. Open the quick-release buckles of the waist belt and chest 

strap.

2. Unfasten both shoulder straps. To do so, push the 

adjusting clamp upwards with your thumb and pull on the 

adjusting clamp.

3. Hang the breathing hoses over the lid.
4. Reach your arms through the shoulder straps and lift up 

the closed-circuit breathing apparatus.

5. Position the closed-circuit breathing apparatus.
6. Make sure that the shoulder straps are tensioned so that 

the padding on the waist belt rests on your hips.

7. Close the waist belt and adjust the width. To do so, pull the 

ends of the belt tight until the closed-circuit breathing 

apparatus rests firmly on your hips.

8. Place the breathing hose in front of your body.
9. Evenly pull the shoulder straps tight and adjust them so 

that the apparatus is comfortable to wear.

10. If necessary, close the chest strap and pull tight.
11. Don the full face mask.

46086

Summary of Contents for BG ProAir

Page 1: ...Instructions for use Dr ger BG ProAir More languages available for download at draeger com ifu Dr ger BG ProAir...

Page 2: ......

Page 3: ...200 0 02 bar h elapsed time 1 4 2 3 27 27 11 16 9 10 8 29 28 24 25 26 14 15 12 13 19 20 23 22 21 5 6 7 17 18 31 30 8 7 6 9 1 2 3 4 5...

Page 4: ...digitalt exemplar p andra spr k i databasen f r teknisk doku mentation www draeger com ifu SV 9300712 ET Selle kasutusjuhendi saab teistes keeltes elektroonili sel kujul alla laadida tehnilise dokume...

Page 5: ...use 12 4 3 1 Freezing the ice packs for apparatuses with ice cooler 12 4 3 2 Inserting ice packs for apparatuses with ice cooler 12 4 3 3 Donning the closed circuit breathing apparatus 12 4 3 4 Activ...

Page 6: ...to potential hazards The meanings of the warning notices are defined as follows 2 2 Trademarks The following website lists the countries in which the trademarks from Dr ger are registered www draeger...

Page 7: ...g cooler options are available Ice cooler The ice cooler contains ice packs that can be refrozen after use Regeneration cooler The regeneration cooler PCM phase chage material cooler contains a latent...

Page 8: ...following information is shown when the compressed oxygen cylinder is opened Other information and queries are shown if the Connect ECU is turned on differently pushbutton insertion of batteries remo...

Page 9: ...mmersion of the apparatus in water e g when navigating a waterhole in a mine will not impair the function of the closed circuit breathing apparatus If the closed circuit breathing apparatus has been c...

Page 10: ...ion on approvals see supplement 9300233 3 6 Type identifying markings and explanation of symbols 3 6 1 Type identifying marking The type identifying marking of the BG ProAir Class 4P is located on the...

Page 11: ...ored for 6 months The apparatus must have successfully passed the leak test and be sealed with the protective cap The air tight shrink wrapped or installed CO2 absorber was not transported on a vehicl...

Page 12: ...ing aid to prevent the ice packs from deforming Make sure that no ice is deposited on the ice packs If necessary place ice packs in a bag seal the bag and freeze the ice packs whilst they remain in th...

Page 13: ...g the automatic distress signal function Remove the function key from the Connect ECU This activates the motion sensor The symbol is displayed 4 4 During use 4 4 1 General information The chemical pro...

Page 14: ...ecess when tightening the bayonet ring Rotate the bayonet ring clockwise onto the connection to fasten the breathing hose The breathing hose is correctly fitted when a clicking sound is heard while ti...

Page 15: ...k the catches on the fasteners under the tabs see Figure B 1 Push the fasteners closed see Figure B 2 Connecting piece The direction of flow is marked with an arrow on the connecting piece The directi...

Page 16: ...g circuit may contain base which may leak when disassembling the closed circuit breathing apparatus and lead to chemical burns Avoid skin contact Wear protective gloves and goggles In case of eye cont...

Page 17: ...th release levers 1 Swivel the spring bridge 90 and remove it from the closed circuit breathing apparatus 2 Release the locking mechanisms between the CO2 absorber and breathing bag as well as between...

Page 18: ...4 Prepare the CO2 absorber Dispose of the disposable absorber For the refill absorber remove the soda lime and the two non woven mats and discard them The sieve can be reused after disinfection 5 2 3...

Page 19: ...rmation 5 2 5 Assembling the closed circuit breathing apparatus 5 2 5 1 Overview 1 Install the CO2 absorber see Installing the CO2 absorber page 19 2 Install the breathing air cooler see Installing th...

Page 20: ...nect the red medium pressure hose to the breathing air cooler The quick connect coupling must engage When connecting make sure that a click is heard If necessary pull on the hose and check whether the...

Page 21: ...th the locking mechanisms to the CO2 absorber and breathing air cooler Every locking mechanism has two lock buttons with engagement hooks Ensure the locking mechanisms on the CO2 absorber and breathin...

Page 22: ...en the Connect ECU on the shoulder strap 5 2 6 Testing the closed circuit breathing apparatus 5 2 6 1 Notes for testing The following test sequence applies for both apparatuses with constant oxygen do...

Page 23: ...he selector switch on the test unit to 2 Tightly pinch the exhalation hose with your hand Use the face spanner from the BG ProAir Test Set if necessary 3 Pump once The test unit must display at least...

Page 24: ...ically turns on and runs a self test 3 Briefly press the bypass button Oxygen must be heard to flow into the circuit flow noise 5 2 6 13 Checking the minimum valve 1 Set the selector switch on the tes...

Page 25: ...b 400 when handling the CO2 absorber WARNING Reduced period of use possible If the expiration date of the soda lime is exceeded the intended period of use for the apparatus may not be reached Note the...

Page 26: ...s are shown on the display 7 Confirm the battery type 8 Turn off the Connect ECU 6 Transport Condensate forms in the breathing circuit during use After use transport the closed circuit breathing appar...

Page 27: ...5 L min Bypass 80 L min Minimum valve 80 L min Temperature during continu ous use1 1 according to EU type examination 6 C to 30 C Air pressure 900 hPa to 1200 hPa Relative humidity 0 100 Storage temp...

Page 28: ...ylinder 2 200 O2 CGA USA R360395 BG ProAir 2 200 O2 Configurator R360560 Accessories Dr ger FPS 7000 RP R56326 Dr ger FPS COM 5000 R62700 klar pilot anti fogging spray R56542 BG ProAir disposable abso...

Page 29: ...tightness test with positive pressure Lower pressure to 7 0 5 mbar Press OK button Pressure difference 1 mbar min Engage lever in spring bridge Pump to activate Opens between 4 and 8 mbar Relief valve...

Page 30: ...e pressure Lower pressure to 7 0 5 mbar Press OK button Pressure difference 1 mbar min Relief valve Engage lever in spring bridge Pump to activate Opens between 4 and 8 mbar Bypass Open cylinder Brief...

Page 31: ...Annex Instructions for use Dr ger BG ProAir 31...

Page 32: ......

Page 33: ......

Page 34: ...9300708H Manufacturer Dr ger Safety AG Co KGaA Revalstra e 1 D 23560 L beck Germany 49 451 8 82 0 9300708 1114 800 en Dr ger Safety AG Co KGaA Edition 01 2021 10 Subject to alterations www draeger com...

Reviews: