Fault / Display (Printer text)
Cause
Remedy
ERROR [XXX]
Instrument defective
(XXX = number specifying the error)
Switch off instrument.
Check whether the error message
reappears after switching on again. If
yes: inform DrägerService. Bring
instrument to repair.
..MAIN-SYSTEM
..IR-SYSTEM
..EC-SYSTEM
..FLOW-SYSTEM
..PRESSURE-SYS
..HEATER-SYSTEM
(Message is printed out additionally)
Instrument blocks automatically all
measurement sequences when error is
detected.
Example:
ERROR [075] HEATER-SYSTEM
The predetermined temperature of the
sensor or the hose is not reached.
Ensure compliance with the ambient
conditions. Switch instrument off,
eventually close the instrument cover
and switch on again.
If the error message reappears after
switching on, inform DrägerService.
23
Fault, Cause and Remedy