6 Maintenance
The device is delivered with default adjustment settings. The date of
the last adjustment can be recalled in the menu.
Dräger recommends that the device be adjusted once a year. Have
the device adjusted only by authorized persons, e.g. DrägerService.
Comply with the national guidelines governing adjustment.
6.1
Charging the rechargeable battery
WARNING
Risk of burns and electric shock
Using the device with a power supply unit that has not been approved
by Dräger may cause burns and electric shocks to the
operator/subject when charging the rechargeable battery.
►
Use only the power supply unit specified in the order list (see
"Order list", page 61).
►
Do not connect the device to the cable and do not charge the
device when carrying out measurements on the subject.
Adjustment interval
expired
The adjustment inter-
val has expired.
Have adjustment car-
ried out by authorized
personnel or Dräger-
Service.
This does not impair
the device being
ready to measure.
Error/Message
Cause
Remedy