This test list can be processed with standard commercially available test aids and tools, but does not replace the inspections and
maintenance work carried out by the manufacturer.
5503.400 Oxylog 1000
Test List
Page 14
K5503400TL1.fm
15.03.06
Co
pyri
ght
reser
ve
d
.
Repro
ducti
on p
e
rmitt
e
d o
n
ly f
o
r non-commerc
ial
purpo
ses.
2.6.3
Testing stenosis alarm
Test set-up:
Fig.
8:
Oxylog 1000 with test set-up: Testing the stenosis alarm
• Set the minute volume adjuster (MV) to 10 L/min.
• Set the frequency selector (Freq.) to 10 1/min.
• Set the Air Mix/No Air Mix selector switch to „Air Mix“.
• Turn the ventilation pressure selector (Paw) to full right.
• During the inhalation phase set a ventilation pressure of approx.
30 mbar (hPa) with the adjustable resistor. (Note: Read off the
ventilation pressure from the pressure gauge of the Oxylog 1000).
• During the inhalation phase slowly close off the ventilation hose further
with the adjustable resistor until the „red“ indicator (Paw high) lights up.
The „Paw high“ switching point is at approx. 55 mbar (hPa). (Read off
switching point from pressure gauge of Oxylog 1000 and from separate
pressure gauge).
O2
Test pressure
reducer
Ventilation hose
Adjustable
resistor
Separate
pressure gauge
T-piece
Summary of Contents for Oxylog 1000
Page 2: ......
Page 5: ...1 General ...
Page 6: ...2 ...
Page 9: ...5 Function Description ...
Page 10: ...6 ...
Page 35: ...31 Maintenance Procedures ...
Page 36: ...32 ...
Page 45: ...41 Schematics and Diagrams ...
Page 46: ...42 ...
Page 55: ...Annex Parts catalog Test List ...
Page 56: ......
Page 58: ...2 ...
Page 64: ...8 ...
Page 65: ...Test List ...
Page 66: ......
Page 103: ......