Function Description
Oxylog 1000
3.4.6
Gas Supply Failure
Warning (Pv)
If the gas supply fails or falls below a defined pressure, the Oxylog 1000
delivers two alarms:
–
Visual alarm indicator "Pv"
–
Audible alarm (alarm generator)
The alarm is triggered under the following conditions:
–
The pressure of the gas supply from the central supply system or from the
O
2
cylinder is inadequate or absent
–
The valve of the O
2
cylinder is closed
–
The gas supply connector is incorrectly installed
Figure 8
Segment of the flow chart: Gas supply failure alarm