Maintenance Procedures
Oxylog 1000
•
Detach the hose from the main switch and connect the pressure gauge
(measuring range: 0 to 6 bar) to the connection socket of the
main
switch
; see following illustration.
•
Screw the compressed gas hose for compressed gas supply onto the
connection socket of the Oxylog 1000.
•
Plug the connector of the compressed gas hose into the wall outlet/com-
pressed gas cylinder.
•
Connect the ventilation hose with flowmeter (measuring range: 10 to
120 L/min) to the connection socket (ventilation hose).
Figure 5
Oxylog 1000 with test set-up: Testing the supply pressure
•
Set the frequency selector (Freq.) to 10 1/min.
•
Set the minute volume adjuster (MV) to full right.
•
Set the ventilation pressure (Paw) to approx. 60 mbar (hPa).
•
Set the Air Mix/No Air Mix selector switch to „No Air Mix“.
•
Set the main switch of the Oxylog 1000 to „I“.
With a flow of approx. 55 L/min (indicated on the flowmeter) the connected
pressure gauge shows a pressure of approx. 1.5 bar.
Flowmeter 10 to
120 L/min
Pressure
gauge 0 to
6 bar
Main switch
Summary of Contents for Oxylog 1000
Page 2: ......
Page 5: ...1 General ...
Page 6: ...2 ...
Page 9: ...5 Function Description ...
Page 10: ...6 ...
Page 35: ...31 Maintenance Procedures ...
Page 36: ...32 ...
Page 45: ...41 Schematics and Diagrams ...
Page 46: ...42 ...
Page 55: ...Annex Parts catalog Test List ...
Page 56: ......
Page 58: ...2 ...
Page 64: ...8 ...
Page 65: ...Test List ...
Page 66: ......
Page 103: ......