This test list can be processed with standard commercially available test aids and tools, but does not replace the inspections and
maintenance work carried out by the manufacturer.
5503.400 Oxylog 1000
Test List
Page 7
K5503400TL1.fm
15.03.06
Co
pyri
ght
reser
ve
d
.
Repro
ducti
on p
e
rmitt
e
d o
n
ly f
o
r non-commerc
ial
purpo
ses.
• Connect the compressed gas connector to the socket of the central
supply/compressed gas cylinder.
• Detach the hose from the main switch and connect the pressure gauge
(measuring range: 0 to 6 bar) to the connection socket of the
main
switch
; see following illustration.
• Connect the ventilation hose with flowmeter (measuring range: 10 to
120 L/min) to the connection socket (ventilation hose).
Fig.
2:
Oxylog 1000 with test set-up: Testing the supply pressure
• Set the frequency selector (Freq.) to 10 1/min.
• Set the minute volume adjuster (MV) to full right.
• Set the ventilation pressure (Paw) to approx. 60 mbar (hPa).
• Set the Air Mix/No Air Mix selector switch to „No Air Mix“.
• Set the main switch of the Oxylog 1000 to „I“.
With a flow of approx. 55 L/min (indicated on the flowmeter) the connected
pressure gauge shows a pressure of approx. 1.5 bar.
• Set the main switch of the Oxylog 1000 to „0“.
Flowmeter 10 to
120 L/min
Pressure
gauge 0 to
6 bar
Main switch
Summary of Contents for Oxylog 1000
Page 2: ......
Page 5: ...1 General ...
Page 6: ...2 ...
Page 9: ...5 Function Description ...
Page 10: ...6 ...
Page 35: ...31 Maintenance Procedures ...
Page 36: ...32 ...
Page 45: ...41 Schematics and Diagrams ...
Page 46: ...42 ...
Page 55: ...Annex Parts catalog Test List ...
Page 56: ......
Page 58: ...2 ...
Page 64: ...8 ...
Page 65: ...Test List ...
Page 66: ......
Page 103: ......