Daily Checkout
3-4
Part Number: 4116841-001
Narkomed MRI-2 Operator’s Manual
Rev: E
WARNING:
Use only nonmagnetic (aluminum) E cylinders with
this machine. Steel cylinders can cause serious injury
or death if brought into an MRI scanning room.
Table 3-1. Nitrous Oxide Cylinder PSI Levels
* typical full load
Pipeline
Supply
System
Verification
11.* Pipeline Supply Verification
a. Inspect the supply hoses for cracks or wear.
b. Connect the appropriate hospital pipeline supply hoses from the
wall outlet fittings to the pipeline inlet connectors.
c.
Check for a sufficient pipeline pressure on the pipeline pressure
gauges located below the flow control valves. The pressure for each
gas should be between 50–55 psi. Open the flow control valve for
each gas to a moderate value. The pressure indicated at the
pipeline pressure gauge must not decrease more than 5 psi.
d. Verify that the correct gases are supplied to the anesthesia machine
inlets.
Low Pressure
System
Verification
12. * Vaporizer Verification
a. Check for a sufficient supply of liquid anesthetic in the vaporizer
using the vaporizer sight glass. The liquid level must be between
the minimum and maximum markings.
b. Make sure the vaporizer fill and drain valves are completely closed.
c.
Check the vaporizer exclusion system, which prevents more than
one vaporizer from being activated simultaneously. Make sure that
when one vaporizer handwheel is turned to a setting greater than
“0”, the other remains locked in its “0” position. Test the system for
both vaporizers. Then, turn all vaporizers to the “0” position.
System Gas
Circuit
Verification
13. * Check the flowmeters. Adjust the flow control knob for each gas and
verify the proper operation of the corresponding flowmeters. The float
must move freely over the full range of each flowmeter.
Oxygen
Monitor
Calibration
14. * Calibrate the oxygen monitor by exposing the sensor to ambient air
and activate the calibration key. (See “Calibrating the Oxygen Sensor”
on page 11-4 for more information.)
GAS
PSI - FULL*
PSI - MIN
Nitrous Oxide
745
600
Oxygen
1900
1000
Summary of Contents for Narkomed MRI-2
Page 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 6: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 10: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 20: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 66: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 74: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 80: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 84: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 88: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 116: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 140: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 172: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 176: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 177: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...